Impact of Ventilator Settings, Nebulizer Placement, Bias Flow, Humidification, and Circuit Adaptor on Aerosol Delivery During Invasive Ventilation

Background: Aerosol delivery during invasive mechanical ventilation is significantly affected by multiple factors related to the patient, device, and ventilator settings. Thus, we aimed to assess the impact of different ventilator settings, nebulizer positions, bias flows, circuit adaptors, and humidification on aerosol delivery during invasive ventilation. Methods: An experimental setup was used to evaluate aerosol delivery using a vibrating mesh nebulizer (VMN) during invasive ventilation (Figure 1). Albuterol at 2.5 mg/3 mL was used under different ventilator settings, nebulizer placements, bias flows, and circuit connectors in the dry circuit. The exhaled humidity and the use of heated humidified circuits were only used in some settings. Each condition was repeated five times. The drug was eluted from the collecting filter and assayed using ultraviolet spectrophotometry at 276 nm. Results: The inhaled dose was higher with high settings compared to low settings regardless of nebulizer position, bias flow, heated humidification, and presence or absence of exhaled humidity (Table 1) (all P < .05) except at 2 L/min bias flow with exhaled humidity and VMN placed proximal to the endotracheal tube in dry circuits. The presence of exhaled humidity or heated humidification tended to diminish the differences between high and low settings in the inhaled dose. In dry circuits, no significant differences were observed in the inhaled dose when comparing Y-shaped and V-shaped connectors, except when the nebulizer was placed proximally at a bias flow of 2 L/min. Also, in dry circuits, proximal placement of the nebulizer at high settings generated higher inhaled doses than distal placement, regardless of the bias flow and connector type used. At low settings, with a Y-connector, no significant differences in the inhaled dose were observed between the two nebulizer placements. Conclusions: In dry circuits, high ventilator settings generated a greater inhaled dose than low settings, with the nebulizer placed 30 cm proximal to the circuit connector showing higher inhaled doses at high settings. Exhaled humidity or the use of heated humidified circuits reduced the differences between high and low settings.Percentage of the inhaled dose delivered during invasive ventilation with high and low ventilator settings in various conditionsNebulizer positionBias flowCircuit connectorPercentage of inhaled dose (%)Dry circuit without exhaled humidityDry circuit with exhaled humidityHeated and humidified circuit with exhaled humidityHigh settingsLow settingsP valueHigh settingsLow settingsP valueHigh settingsLow settingsP valueProximal2V36.9±1.620.7±1.8.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y39.8±1.718.3±1.6.00925.7±4.320.1±1.7.076&nbsp;&nbsp;&nbsp;P.021.076&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;5V32.8±1.516.6±1.8.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y34.5±0.816.3±1.4.00921.8±1.117.7±1.3.00923.3±1.219.3±1.9.009P.076.917&nbsp;.117.076&nbsp;&nbsp;&nbsp;&nbsp;Distal2V31.0±2.415.4±1.6.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y31.1±1.618.2±2.3.00923.0±1.617.2±1.3.009&nbsp;&nbsp;&nbsp;P.175.076&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;5V31.7±2.118.3±1.6.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y28.7±2.317.9±2.0.00923.7±2.216.9±1.6.00926.1±2.419.5±1.7.009P.754.754&nbsp;.602.675&nbsp;.076.917&nbsp;High settings: VT: 500mL, RR:12bpm, Ti: 2.5 sec, Tplat: 0.5 sec, PEEP: 5 cmH2O; Low settings: VT: 380mL, RR:16bpm, Ti: 1.2 sec, Tplat: 0 sec, and PEEP: 5 cmH2O; Bias flow reported in L/min; Proximal: VMN placed 30 cm away from the circuit connection; Distal: VMN placed at the humidifier inlet with heated and humidified circuit or at the ventilator outlet with dry circuit.A) The study set up of a simulated adult intubated with an endotracheal tube (ETT) and mechanically ventilated while receiving invasive aerosol delivery using a vibrating mesh nebulizer (VMN). B) proximal VMN position placed 30 cm away from the circuit connection. C) distal VMN position placed at the humidifier inlet with the heated and humidified circuit. D) distal VMN position placed at the ventilator outlet with the dry circuit. E) Y-shaped connector. F) V-shaped connector.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2024-10-01
DOI
https://doi.org/10.4187/respcare.20244114853
Primary Topic
Inhalation and Respiratory Drug Delivery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of Ventilator Settings, Nebulizer Placement, Bias Flow, Humidification, and Circuit Adaptor on Aerosol Delivery During Invasive Ventilation

Omar Alamoudi, Fai Ali Albuainain, Xiaoyan Man, Jie Li
Respiratory Care
Inhalation and Respiratory Drug Delivery
article

Impact of Ventilator Settings, Nebulizer Placement, Bias Flow, Humidification, and Circuit Adaptor on Aerosol Delivery During Invasive Ventilation

Omar Alamoudi, Fai Ali Albuainain, Xiaoyan Man, Jie Li
article en

Abstract

Background: Aerosol delivery during invasive mechanical ventilation is significantly affected by multiple factors related to the patient, device, and ventilator settings. Thus, we aimed to assess the impact of different ventilator settings, nebulizer positions, bias flows, circuit adaptors, and humidification on aerosol delivery during invasive ventilation. Methods: An experimental setup was used to evaluate aerosol delivery using a vibrating mesh nebulizer (VMN) during invasive ventilation (Figure 1). Albuterol at 2.5 mg/3 mL was used under different ventilator settings, nebulizer placements, bias flows, and circuit connectors in the dry circuit. The exhaled humidity and the use of heated humidified circuits were only used in some settings. Each condition was repeated five times. The drug was eluted from the collecting filter and assayed using ultraviolet spectrophotometry at 276 nm. Results: The inhaled dose was higher with high settings compared to low settings regardless of nebulizer position, bias flow, heated humidification, and presence or absence of exhaled humidity (Table 1) (all P < .05) except at 2 L/min bias flow with exhaled humidity and VMN placed proximal to the endotracheal tube in dry circuits. The presence of exhaled humidity or heated humidification tended to diminish the differences between high and low settings in the inhaled dose. In dry circuits, no significant differences were observed in the inhaled dose when comparing Y-shaped and V-shaped connectors, except when the nebulizer was placed proximally at a bias flow of 2 L/min. Also, in dry circuits, proximal placement of the nebulizer at high settings generated higher inhaled doses than distal placement, regardless of the bias flow and connector type used. At low settings, with a Y-connector, no significant differences in the inhaled dose were observed between the two nebulizer placements. Conclusions: In dry circuits, high ventilator settings generated a greater inhaled dose than low settings, with the nebulizer placed 30 cm proximal to the circuit connector showing higher inhaled doses at high settings. Exhaled humidity or the use of heated humidified circuits reduced the differences between high and low settings.Percentage of the inhaled dose delivered during invasive ventilation with high and low ventilator settings in various conditionsNebulizer positionBias flowCircuit connectorPercentage of inhaled dose (%)Dry circuit without exhaled humidityDry circuit with exhaled humidityHeated and humidified circuit with exhaled humidityHigh settingsLow settingsP valueHigh settingsLow settingsP valueHigh settingsLow settingsP valueProximal2V36.9±1.620.7±1.8.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y39.8±1.718.3±1.6.00925.7±4.320.1±1.7.076&nbsp;&nbsp;&nbsp;P.021.076&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;5V32.8±1.516.6±1.8.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y34.5±0.816.3±1.4.00921.8±1.117.7±1.3.00923.3±1.219.3±1.9.009P.076.917&nbsp;.117.076&nbsp;&nbsp;&nbsp;&nbsp;Distal2V31.0±2.415.4±1.6.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y31.1±1.618.2±2.3.00923.0±1.617.2±1.3.009&nbsp;&nbsp;&nbsp;P.175.076&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;5V31.7±2.118.3±1.6.009&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;&nbsp;Y28.7±2.317.9±2.0.00923.7±2.216.9±1.6.00926.1±2.419.5±1.7.009P.754.754&nbsp;.602.675&nbsp;.076.917&nbsp;High settings: VT: 500mL, RR:12bpm, Ti: 2.5 sec, Tplat: 0.5 sec, PEEP: 5 cmH2O; Low settings: VT: 380mL, RR:16bpm, Ti: 1.2 sec, Tplat: 0 sec, and PEEP: 5 cmH2O; Bias flow reported in L/min; Proximal: VMN placed 30 cm away from the circuit connection; Distal: VMN placed at the humidifier inlet with heated and humidified circuit or at the ventilator outlet with dry circuit.A) The study set up of a simulated adult intubated with an endotracheal tube (ETT) and mechanically ventilated while receiving invasive aerosol delivery using a vibrating mesh nebulizer (VMN). B) proximal VMN position placed 30 cm away from the circuit connection. C) distal VMN position placed at the humidifier inlet with the heated and humidified circuit. D) distal VMN position placed at the ventilator outlet with the dry circuit. E) Y-shaped connector. F) V-shaped connector.

Respiratory CareVol. 69(10_suppl)
Illinois College (US), University of Illinois Chicago (US), Rush University (US), Imam Abdulrahman Bin Faisal University (SA)
Openalex Percentile: Top 41%
Inhalation and Respiratory Drug Delivery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.