Developing a 3D Printed Fluidic Positive Airway Pressure Device: A Bench Comparison

Background: Positive airway pressure (PAP) falls in the category of hyperinflation therapy. Hyperinflation therapy is used to prevent or reverse atelectasis, reduce air trapping in obstructive diseases, aids in mobilization of retained secretions and/or optimizes inhaled drug delivery. PAP devices are separated from positive expiratory pressure (PEP) devices, due to the addition of flow that adds inspiratory as well as expiratory positive pressure. There are two types of PAP devices on the market today: threshold and flow resistors. The threshold resistors are more accurate regardless of flow due to a spring-loaded valve but are more expensive to manufacturer. Flow resistors are less precise due to varying inspiratory and expiratory flows of the patient but cheaper to manufacturer. The UTMB CRAVE lab in partnership with FluidIQ, Inc with their combined knowledge of fluidics set out to create a more accurate flow resistor PAP device, we labeled FPAP. Methods: Two commercially available PAP devices (AccuPAP and VersaPAP) were compared to three of the best versions of FPAP (V13, V14, V15) 3D printed in the CRAVE lab. All devices were tested using the ASL 5000 (IngMar) in two patient conditions: low (PIF 21.4, PEF 12.7 L/min) and high flow (PIF 53.9, PEF 48.4) breathing characteristics. All devices were set to achieve 5, 10, 15, and 20 cm H2O and flow recorded on a calibrated flow meter (Fluke) prior to experiment. Each device was tested for 50 breaths (25 low and 25 high flow). ± 2 cm H2O of desired target PAP was considered significant. Results: Table 1 provides a summary of the results. Using commercially available devices as the standard, V13 (62.5%) fell below, while V15 (75%) matched and V14 (78.6%) exceeded the % acceptable PAP delivery. Conclusions: 3D printing allows for the rapid prototyping and testing of new PAP designs aimed at improving efficiency (gas consumption) and precision of delivered PAP. Through this novel design we were able to produce 2 devices that match or exceed currently available products PAP delivery in a simulated model.Table 1.

Authors

Institutions

Publication Details

Journal
Respiratory Care
Published
2024-10-01
DOI
https://doi.org/10.4187/respcare.20244137680
Primary Topic
Inhalation and Respiratory Drug Delivery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Developing a 3D Printed Fluidic Positive Airway Pressure Device: A Bench Comparison

Brian K Walsh, Duncan Stevenson, Khamron Micheals
Respiratory Care
Inhalation and Respiratory Drug Delivery
article

Developing a 3D Printed Fluidic Positive Airway Pressure Device: A Bench Comparison

Brian K Walsh, Duncan Stevenson, Khamron Micheals
article en

Abstract

Background: Positive airway pressure (PAP) falls in the category of hyperinflation therapy. Hyperinflation therapy is used to prevent or reverse atelectasis, reduce air trapping in obstructive diseases, aids in mobilization of retained secretions and/or optimizes inhaled drug delivery. PAP devices are separated from positive expiratory pressure (PEP) devices, due to the addition of flow that adds inspiratory as well as expiratory positive pressure. There are two types of PAP devices on the market today: threshold and flow resistors. The threshold resistors are more accurate regardless of flow due to a spring-loaded valve but are more expensive to manufacturer. Flow resistors are less precise due to varying inspiratory and expiratory flows of the patient but cheaper to manufacturer. The UTMB CRAVE lab in partnership with FluidIQ, Inc with their combined knowledge of fluidics set out to create a more accurate flow resistor PAP device, we labeled FPAP. Methods: Two commercially available PAP devices (AccuPAP and VersaPAP) were compared to three of the best versions of FPAP (V13, V14, V15) 3D printed in the CRAVE lab. All devices were tested using the ASL 5000 (IngMar) in two patient conditions: low (PIF 21.4, PEF 12.7 L/min) and high flow (PIF 53.9, PEF 48.4) breathing characteristics. All devices were set to achieve 5, 10, 15, and 20 cm H2O and flow recorded on a calibrated flow meter (Fluke) prior to experiment. Each device was tested for 50 breaths (25 low and 25 high flow). ± 2 cm H2O of desired target PAP was considered significant. Results: Table 1 provides a summary of the results. Using commercially available devices as the standard, V13 (62.5%) fell below, while V15 (75%) matched and V14 (78.6%) exceeded the % acceptable PAP delivery. Conclusions: 3D printing allows for the rapid prototyping and testing of new PAP designs aimed at improving efficiency (gas consumption) and precision of delivered PAP. Through this novel design we were able to produce 2 devices that match or exceed currently available products PAP delivery in a simulated model.Table 1.

Respiratory CareVol. 69(10_suppl)
Texas Medical Center (US), Dell Children's Medical Center of Central Texas (US), The University of Texas Medical Branch at Galveston (US), Fluidigm (United States) (US)
Openalex Percentile: Top 41%
Inhalation and Respiratory Drug Delivery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.