Technical development and implementation of 3D‐QALAS on a 1.5T MR‐Linac for the brain: A prospective R‐IDEAL stage 0/1 technology development report

BACKGROUND: Quantitative relaxometry of the brain on the integrated MRI / linear accelerator (MR-Linac) at high isotropic reconstructed resolution is currently limited due to prohibitively long scan times and limited field-of-views. PURPOSE: The purpose of this study was to assess the technical feasibility of the 3D-QALAS technique on the 1.5T MR-Linac which can acquire whole-brain 1 mm isotropic quantitative T1, T2, and PD maps along with multiple synthetic images in a 7-minute acquisition time. METHODS: A 1 mm isotropic 3D-QALAS acquisition was scanned in both phantoms and five healthy volunteers on the 1.5T Elekta Unity MR-Linac device with scan times around seven minutes. A test-retest protocol across five independent sessions for the phantom was conducted. The Lin's Concordance Correlation Coefficient (LCCC), repeatability, and reproducibility between measured and reference quantitative T1, T2, and PD values were determined in the phantom. Distortion was also studied. Vendor-provided reconstruction through SyMRI was performed to extract synthetic images and brain volume metric assessments in the healthy volunteers. RESULTS: The slope and LCCC between the measured and phantom reference values was 1.02 (1.00), 1.09 (0.90), and 0.99 (1.00) for T1, T2, and PD, respectively. Median distortion across the phantom remained below 2 mm. The repeatability and reproducibility coefficient-of-variation (CoV) was under 8% for all measured values. The measured brain volumes in the healthy volunteer was within expected age-adjusted reference values. CONCLUSIONS: The technical feasibility of using 3D-QALAS on the integrated 1.5T MR-Linac was confirmed. Applying this technique to the adaptive radiation therapy workflow of the brain will provide new opportunities to integrate quantitative imaging relaxometry biomarkers at 1 mm isotropic reconstructed resolution.

Authors

Institutions

Publication Details

Journal
Journal of Applied Clinical Medical Physics
Published
2026-10-28
DOI
https://doi.org/10.1002/acm2.70886
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Technical development and implementation of 3D‐QALAS on a 1.5T MR‐Linac for the brain: A prospective R‐IDEAL stage 0/1 technology development report

Ken‐Pin Hwang, Natalie A. West, Ergys Subashi, Brian A. Taylor et al.
Journal of Applied Clinical Medical Physics
Advanced MRI Techniques and Applications
article

Technical development and implementation of 3D‐QALAS on a 1.5T MR‐Linac for the brain: A prospective R‐IDEAL stage 0/1 technology development report

Ken‐Pin Hwang, Natalie A. West, Ergys Subashi, Brian A. Taylor, Clifton David Fuller, Lucas B. McCullum, Ashley Harrington, Clifton D. Fuller, Grace Nehring
article en

Abstract

BACKGROUND: Quantitative relaxometry of the brain on the integrated MRI / linear accelerator (MR-Linac) at high isotropic reconstructed resolution is currently limited due to prohibitively long scan times and limited field-of-views. PURPOSE: The purpose of this study was to assess the technical feasibility of the 3D-QALAS technique on the 1.5T MR-Linac which can acquire whole-brain 1 mm isotropic quantitative T1, T2, and PD maps along with multiple synthetic images in a 7-minute acquisition time. METHODS: A 1 mm isotropic 3D-QALAS acquisition was scanned in both phantoms and five healthy volunteers on the 1.5T Elekta Unity MR-Linac device with scan times around seven minutes. A test-retest protocol across five independent sessions for the phantom was conducted. The Lin's Concordance Correlation Coefficient (LCCC), repeatability, and reproducibility between measured and reference quantitative T1, T2, and PD values were determined in the phantom. Distortion was also studied. Vendor-provided reconstruction through SyMRI was performed to extract synthetic images and brain volume metric assessments in the healthy volunteers. RESULTS: The slope and LCCC between the measured and phantom reference values was 1.02 (1.00), 1.09 (0.90), and 0.99 (1.00) for T1, T2, and PD, respectively. Median distortion across the phantom remained below 2 mm. The repeatability and reproducibility coefficient-of-variation (CoV) was under 8% for all measured values. The measured brain volumes in the healthy volunteer was within expected age-adjusted reference values. CONCLUSIONS: The technical feasibility of using 3D-QALAS on the integrated 1.5T MR-Linac was confirmed. Applying this technique to the adaptive radiation therapy workflow of the brain will provide new opportunities to integrate quantitative imaging relaxometry biomarkers at 1 mm isotropic reconstructed resolution.

Journal of Applied Clinical Medical PhysicsVol. 27(11)
The University of Texas MD Anderson Cancer Center (US), The University of Texas Health Science Center at Houston (US)
National Science Foundation, Elekta, NRG Oncology, Siemens Healthineers, National Institutes of Health, University of Texas MD Anderson Cancer Center, National Institute on Drug Abuse, National Cancer Institute, National Institute of Dental and Craniofacial Research
Good health and well-being
Openalex Percentile: Top 88%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon