Absolute scaling of small- and wide-angle X-ray scattering images recorded with short duration X-ray pulses on a large area fiber-taper X-ray detector

We have developed an X-ray scattering setup capable of capturing time-resolved small-angle X-ray scattering (SAXS) and wide-angle X-ray scattering (WAXS) images spanning q = 0.02 to over 5.2 Å<sup>-1</sup> on a single, large area Rayonix MX340-HS detector with time resolution as short as 120 ps. A key feature of this setup is a 0.51 mm-diameter partially transmissive beamstop that enables non-invasive, image-by-image recording of direct beam position and intensity during acquisition. This beamstop attenuates 12 keV undulator radiation by approximately eight orders of magnitude while suppressing off-axis second harmonic radiation. Continuous monitoring of the direct beam position facilitates long-term beam alignment and allows datasets acquired at different times to be placed on a common absolute scale prior to differencing. The accuracy of the difference scattering curves is ultimately limited by the performance of the large area, fiber-taper X-ray detector used in this study. To address accuracy issues, we present a detailed statistical characterization of the detector readout noise and responsivity and introduce a variance-per-count statistic that enables identification of zinger-free averages, generation of precise uniformity corrections, and statistically weighted conversion of two-dimensional scattering images into near shot-noise-limited one-dimensional scattering curves. The detector point-spread function and its effects on resolution and scaling are examined using scattering data from a fused silica plate and from apoferritin solution in a capillary. The ability to acquire high accuracy, high precision scattering curves over a broad range of q and temperatures provides a robust foundation for time-resolved studies of biomolecular structure and dynamics in solution.

Authors

Institutions

Publication Details

Journal
Journal of Synchrotron Radiation
Published
2026-02-26
DOI
https://doi.org/10.1107/s1600577526001451
Primary Topic
Enzyme Structure and Function
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Absolute scaling of small- and wide-angle X-ray scattering images recorded with short duration X-ray pulses on a large area fiber-taper X-ray detector

Friedrich Schotte, Philip Anfinrud, H. S. Cho, Robert Henning
Journal of Synchrotron Radiation
Enzyme Structure and Function
article

Absolute scaling of small- and wide-angle X-ray scattering images recorded with short duration X-ray pulses on a large area fiber-taper X-ray detector

Friedrich Schotte, Philip Anfinrud, H. S. Cho, Robert Henning
article en

Abstract

We have developed an X-ray scattering setup capable of capturing time-resolved small-angle X-ray scattering (SAXS) and wide-angle X-ray scattering (WAXS) images spanning q = 0.02 to over 5.2 Å<sup>-1</sup> on a single, large area Rayonix MX340-HS detector with time resolution as short as 120 ps. A key feature of this setup is a 0.51 mm-diameter partially transmissive beamstop that enables non-invasive, image-by-image recording of direct beam position and intensity during acquisition. This beamstop attenuates 12 keV undulator radiation by approximately eight orders of magnitude while suppressing off-axis second harmonic radiation. Continuous monitoring of the direct beam position facilitates long-term beam alignment and allows datasets acquired at different times to be placed on a common absolute scale prior to differencing. The accuracy of the difference scattering curves is ultimately limited by the performance of the large area, fiber-taper X-ray detector used in this study. To address accuracy issues, we present a detailed statistical characterization of the detector readout noise and responsivity and introduce a variance-per-count statistic that enables identification of zinger-free averages, generation of precise uniformity corrections, and statistically weighted conversion of two-dimensional scattering images into near shot-noise-limited one-dimensional scattering curves. The detector point-spread function and its effects on resolution and scaling are examined using scattering data from a fused silica plate and from apoferritin solution in a capillary. The ability to acquire high accuracy, high precision scattering curves over a broad range of q and temperatures provides a robust foundation for time-resolved studies of biomolecular structure and dynamics in solution.

Journal of Synchrotron RadiationVol. 33(3)
National Institutes of Health (US), University of Chicago (US)
Openalex Percentile: Top 28%
Enzyme Structure and Function
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.