Intraoperative TEE Detection of Guidewire Malposition and Incomplete Valve Expansion in Valve-in-Valve TAVI for Trifecta Bioprosthesis Failure: a Case Report

Structured Abstract Background Valve-in-valve transcatheter aortic valve implantation (ViV-TAVI) is widely used for failed surgical bioprostheses; however, procedural pitfalls remain, particularly in valves prone to early structural deterioration. Case summary An 84-year-old woman with a failed 21-mm Trifecta bioprosthesis underwent ViV-TAVI. Although the guidewire crossed smoothly and fluoroscopy appeared reassuring, incomplete valve expansion occurred. Intraoperative transoesophageal echocardiography revealed that the guidewire had passed between degenerated leaflets rather than through the true central orifice. After repositioning under TEE guidance, full symmetric valve expansion and optimal haemodynamics were achieved. Discussion This case highlights the risk of false procedural reassurance during ViV-TAVI and underscores the importance of actively confirming central guidewire trajectory, particularly in failed Trifecta bioprostheses.

Authors

Institutions

Publication Details

Journal
European Heart Journal - Case Reports
Published
2026-03-08
DOI
https://doi.org/10.1093/ehjcr/ytag183
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intraoperative TEE Detection of Guidewire Malposition and Incomplete Valve Expansion in Valve-in-Valve TAVI for Trifecta Bioprosthesis Failure: a Case Report

Hirokuni Akahori, Taichi Sakaguchi, Nagataka Yoshihara, Kenichi Watanabe
European Heart Journal - Case Reports
Cardiac Valve Diseases and Treatments
article

Intraoperative TEE Detection of Guidewire Malposition and Incomplete Valve Expansion in Valve-in-Valve TAVI for Trifecta Bioprosthesis Failure: a Case Report

Hirokuni Akahori, Taichi Sakaguchi, Nagataka Yoshihara, Kenichi Watanabe
article en

Abstract

Structured Abstract Background Valve-in-valve transcatheter aortic valve implantation (ViV-TAVI) is widely used for failed surgical bioprostheses; however, procedural pitfalls remain, particularly in valves prone to early structural deterioration. Case summary An 84-year-old woman with a failed 21-mm Trifecta bioprosthesis underwent ViV-TAVI. Although the guidewire crossed smoothly and fluoroscopy appeared reassuring, incomplete valve expansion occurred. Intraoperative transoesophageal echocardiography revealed that the guidewire had passed between degenerated leaflets rather than through the true central orifice. After repositioning under TEE guidance, full symmetric valve expansion and optimal haemodynamics were achieved. Discussion This case highlights the risk of false procedural reassurance during ViV-TAVI and underscores the importance of actively confirming central guidewire trajectory, particularly in failed Trifecta bioprostheses.

European Heart Journal - Case Reports
Hyogo University (JP)
Gender equality
Openalex Percentile: Top 34%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Intraoperative TEE Detection of Guidewire Malposition and Incomplete Valve Expansion in Valve-in-Valve TAVI for Trifecta Bioprosthesis Failure: a Case Report — Hirokuni Akahori, Taichi Sakaguchi, et al. · European Heart Journal - Case Reports (2026) | TGRS Research Map | TGRS