To evaluate the technical feasibility of a standardized, sclerotomy-free method for injecting viscoelastic material into the suprachoroidal space using a novel mechanical injector, and to evaluate an upgraded injector version in an ex vivo porcine model. This laboratory investigation was conducted at a central medical research center using thirteen freshly enucleated pig eyes obtained from a local farm. All eyes underwent a 360-degree peritomy and were fixed to a Styrofoam platform. In cohort 1 (n = 5), a custom mechanical injector (Supra SEG) with a 30-gauge needle and selectable penetration depths (0.7 mm and 0.9 mm) was used to inject saline or viscoelastic materials (Biolon Prime, Healon 5 Pro, and Healon EndoCoat) into the suprachoroidal space (SCS). In cohort 2 (n = 8), an upgraded injector (27-gauge needle and larger-diameter tubing) was evaluated using the 0.9 mm depth. Ultrasound biomicroscopy (UBM) was performed before and after each injection to confirm SCS localization. The primary outcome was successful UBM-verified delivery into the SCS. In cohort 1, injections at 0.7 mm were unsuccessful in all attempts, whereas 0.9 mm enabled correct SCS delivery in all eyes. In cohort 2, successful UBM-confirmed suprachoroidal delivery was achieved in all eight eyes across multiple viscoelastic materials, including sequential and repeat injections that produced a larger, merged suprachoroidal bleb. UBM demonstrated characteristic echogenic appearances for each material. Injection resistance in cohort 2 was consistently mild to moderate. A sclerotomy-free, image-guided, tangential suprachoroidal injection technique using a purpose-built injector is technically feasible in ex vivo porcine eyes. Because this post-mortem model cannot adequately assess hemorrhagic complications, procedural safety and clinical performance require further in vivo evaluation.

Authors

Institutions

Publication Details

Journal
International Journal of Retina and Vitreous
Published
2026-03-11
DOI
https://doi.org/10.1186/s40942-026-00831-4
Primary Topic
Retinal and Macular Surgery
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Simplified suprachoroidal buckling: sclerotomy-free viscoelastic injection with a novel device in porcine eyes

Assaf Dotan, Roee Arnon, Yariv Keshet, Orly Gal-Or et al.
International Journal of Retina and Vitreous
Retinal and Macular Surgery
article

Simplified suprachoroidal buckling: sclerotomy-free viscoelastic injection with a novel device in porcine eyes

Assaf Dotan, Roee Arnon, Yariv Keshet, Orly Gal-Or, Assaf Ben-Arzi
article en

Abstract

To evaluate the technical feasibility of a standardized, sclerotomy-free method for injecting viscoelastic material into the suprachoroidal space using a novel mechanical injector, and to evaluate an upgraded injector version in an ex vivo porcine model. This laboratory investigation was conducted at a central medical research center using thirteen freshly enucleated pig eyes obtained from a local farm. All eyes underwent a 360-degree peritomy and were fixed to a Styrofoam platform. In cohort 1 (n = 5), a custom mechanical injector (Supra SEG) with a 30-gauge needle and selectable penetration depths (0.7 mm and 0.9 mm) was used to inject saline or viscoelastic materials (Biolon Prime, Healon 5 Pro, and Healon EndoCoat) into the suprachoroidal space (SCS). In cohort 2 (n = 8), an upgraded injector (27-gauge needle and larger-diameter tubing) was evaluated using the 0.9 mm depth. Ultrasound biomicroscopy (UBM) was performed before and after each injection to confirm SCS localization. The primary outcome was successful UBM-verified delivery into the SCS. In cohort 1, injections at 0.7 mm were unsuccessful in all attempts, whereas 0.9 mm enabled correct SCS delivery in all eyes. In cohort 2, successful UBM-confirmed suprachoroidal delivery was achieved in all eight eyes across multiple viscoelastic materials, including sequential and repeat injections that produced a larger, merged suprachoroidal bleb. UBM demonstrated characteristic echogenic appearances for each material. Injection resistance in cohort 2 was consistently mild to moderate. A sclerotomy-free, image-guided, tangential suprachoroidal injection technique using a purpose-built injector is technically feasible in ex vivo porcine eyes. Because this post-mortem model cannot adequately assess hemorrhagic complications, procedural safety and clinical performance require further in vivo evaluation.

International Journal of Retina and Vitreous
Tel Aviv University (IL), Rabin Medical Center (IL)
Openalex Percentile: Top 28%
Retinal and Macular Surgery
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.