Influence of attachment designs in the insertion and removal force of clear aligners: An experimental study of strain sensors

Objectives: This experimental study analyzed the influence of attachment designs on the insertion and removal force of orthodontic clear aligners (CA). Material and Methods: Eleven CA (0.63 mm thickness) were manufactured with rectangle, beveled, distal slice, round wedge, pyramidal, and half-moon attachment shapes. The attachments were included in dental units 11, 12, 13, and 16, and absence of attachments in the control group. Analyses of the deformation stress (-) and expansion (+) of the CA were recorded six times using a strain gage during appliance insertion and removal. Descriptive analyses were calculated for all the study variables, followed by the analysis of variance test for repeated measures (significance level of 0.05). Results: Despite no attachments, CA still had low expansion (2.730; Standard deviation [SD] = 0.327) alteration. Rectangle horizontal scale in the dental unit 16 had the highest strain (−13.869; SD = 1.198 µm/m), similar to the half-moon type (−13.009; SD = 2.399 µm/m) in dental unit 11. CA with a beveled horizontal scale in dental unit 16 had a higher expansion threshold (5.337; SD = 0.453 µm/m) than the beveled low-profile horizontal scale in dental unit 12 (1.046; SD = 1.010 µm/m). Conclusion: Attachments significantly influence this behavior, with certain designs, such as the half-moon and rectangular types, producing higher stress and expansion values. The control group showed minimal strain, confirming the attachments’ role in guiding force distribution. These findings highlight the importance of attachment design in optimizing aligner performance and treatment effectiveness.

Authors

Institutions

Publication Details

Journal
APOS Trends in Orthodontics
Published
2026-03-11
DOI
https://doi.org/10.25259/apos_202_2025
Primary Topic
Orthodontics and Dentofacial Orthopedics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of attachment designs in the insertion and removal force of clear aligners: An experimental study of strain sensors

Carlo Marassi, Lincoln Issamu Nojima, Bruna Caroline Tomé Barreto, Luísa Schubach da Costa Barreto et al.
APOS Trends in Orthodontics
Orthodontics and Dentofacial Orthopedics
article

Influence of attachment designs in the insertion and removal force of clear aligners: An experimental study of strain sensors

Carlo Marassi, Lincoln Issamu Nojima, Bruna Caroline Tomé Barreto, Luísa Schubach da Costa Barreto, Guido Artemio Marañón-Vásquez, Matilde da Cunha Gonçalves NOJIMA, Ana Luisa Fernandes Santos Raposo Viana, Carlos Nelson Elias
article en

Abstract

Objectives: This experimental study analyzed the influence of attachment designs on the insertion and removal force of orthodontic clear aligners (CA). Material and Methods: Eleven CA (0.63 mm thickness) were manufactured with rectangle, beveled, distal slice, round wedge, pyramidal, and half-moon attachment shapes. The attachments were included in dental units 11, 12, 13, and 16, and absence of attachments in the control group. Analyses of the deformation stress (-) and expansion (+) of the CA were recorded six times using a strain gage during appliance insertion and removal. Descriptive analyses were calculated for all the study variables, followed by the analysis of variance test for repeated measures (significance level of 0.05). Results: Despite no attachments, CA still had low expansion (2.730; Standard deviation [SD] = 0.327) alteration. Rectangle horizontal scale in the dental unit 16 had the highest strain (−13.869; SD = 1.198 µm/m), similar to the half-moon type (−13.009; SD = 2.399 µm/m) in dental unit 11. CA with a beveled horizontal scale in dental unit 16 had a higher expansion threshold (5.337; SD = 0.453 µm/m) than the beveled low-profile horizontal scale in dental unit 12 (1.046; SD = 1.010 µm/m). Conclusion: Attachments significantly influence this behavior, with certain designs, such as the half-moon and rectangular types, producing higher stress and expansion values. The control group showed minimal strain, confirming the attachments’ role in guiding force distribution. These findings highlight the importance of attachment design in optimizing aligner performance and treatment effectiveness.

APOS Trends in OrthodonticsVol. 0
Universidade Federal do Rio de Janeiro (BR), Universidade de Ribeirão Preto (BR), Universidade do Estado do Rio de Janeiro (BR), Military Institute of Engineering (BR)
Openalex Percentile: Top 27%
Orthodontics and Dentofacial Orthopedics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.