Thermoacoustic Ultrasound Assessment of Liver Steatosis—A Novel Approach for MASLD Diagnosis

<b>Background/Objectives</b>: Metabolic Dysfunction-Associated Steatotic Liver Disease (MASLD) is a global health crisis, but current diagnostics are limited. Liver biopsy is invasive, magnetic resonance imaging-proton density fat fraction (MRI-PDFF) is expensive, and quantitative ultrasound methods are low-accuracy, especially in patients with a high body mass index (BMI). This study introduces a novel thermo-acoustic (TA) method that generates ultrasound signals based on tissue electrical conductivity, where lean tissue (high in water and electrolytes) absorbs more radio-frequency (RF) energy than fatty tissue, providing a direct molecular contrast for fat. <b>Methods</b>: A prospective, cross-sectional feasibility study compared a new thermo-acoustic fat fraction (TAFF) score with the reference standard MRI-PDFF in 40 subjects with suspected fatty liver disease. Bland-Altman analysis, Deming regression, and Binary classification performance were tested. To establish system stability, a dedicated Repeatability and Reproducibility (R&R) study (N = 14) evaluated inter-operator and intra-operator consistency using an Intraclass Correlation Coefficient (ICC) derived from a two-way random-effects ANOVA model. <b>Results:</b> TAFF estimates demonstrated a substantial correlation (r = 0.89) with MRI-PDFF and an average absolute error of 3.04% fat fraction. Classification performance was high, with an Area Under the Receiver Operating Characteristic Curve (AUROC) of 0.92 at the 12% fat fraction threshold and 0.99 at the 20% fat fraction threshold. The R&R study confirmed robust stability (intraclass correlation = 0.89) and a negligible mean inter-operator difference of 0.36%. Estimation errors showed no statistically significant correlation with BMI or other body habitus measurements. <b>Conclusions</b>: These findings support thermoacoustics' potential as an accurate, non-invasive, point-of-care solution that can serve as a new imaging biomarker. By providing predictive values closely aligned with MRI-PDFF across the full MASLD spectrum, TAFF can complement currently available ultrasound methods to address the cost and access constraints of MRI for the assessment, diagnosis, and monitoring of MASLD.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2026-03-09
DOI
https://doi.org/10.3390/diagnostics16050804
Primary Topic
Liver Disease Diagnosis and Treatment
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Thermoacoustic Ultrasound Assessment of Liver Steatosis—A Novel Approach for MASLD Diagnosis

Idan Steinberg, Michael Thornton, Jonathan M. Rubin, Jing Gao et al.
Diagnostics
Liver Disease Diagnosis and Treatment
article

Thermoacoustic Ultrasound Assessment of Liver Steatosis—A Novel Approach for MASLD Diagnosis

Idan Steinberg, Michael Thornton, Jonathan M. Rubin, Jing Gao, Jang Hwan Cho, Christopher M. Bull
article en

Abstract

<b>Background/Objectives</b>: Metabolic Dysfunction-Associated Steatotic Liver Disease (MASLD) is a global health crisis, but current diagnostics are limited. Liver biopsy is invasive, magnetic resonance imaging-proton density fat fraction (MRI-PDFF) is expensive, and quantitative ultrasound methods are low-accuracy, especially in patients with a high body mass index (BMI). This study introduces a novel thermo-acoustic (TA) method that generates ultrasound signals based on tissue electrical conductivity, where lean tissue (high in water and electrolytes) absorbs more radio-frequency (RF) energy than fatty tissue, providing a direct molecular contrast for fat. <b>Methods</b>: A prospective, cross-sectional feasibility study compared a new thermo-acoustic fat fraction (TAFF) score with the reference standard MRI-PDFF in 40 subjects with suspected fatty liver disease. Bland-Altman analysis, Deming regression, and Binary classification performance were tested. To establish system stability, a dedicated Repeatability and Reproducibility (R&R) study (N = 14) evaluated inter-operator and intra-operator consistency using an Intraclass Correlation Coefficient (ICC) derived from a two-way random-effects ANOVA model. <b>Results:</b> TAFF estimates demonstrated a substantial correlation (r = 0.89) with MRI-PDFF and an average absolute error of 3.04% fat fraction. Classification performance was high, with an Area Under the Receiver Operating Characteristic Curve (AUROC) of 0.92 at the 12% fat fraction threshold and 0.99 at the 20% fat fraction threshold. The R&R study confirmed robust stability (intraclass correlation = 0.89) and a negligible mean inter-operator difference of 0.36%. Estimation errors showed no statistically significant correlation with BMI or other body habitus measurements. <b>Conclusions</b>: These findings support thermoacoustics' potential as an accurate, non-invasive, point-of-care solution that can serve as a new imaging biomarker. By providing predictive values closely aligned with MRI-PDFF across the full MASLD spectrum, TAFF can complement currently available ultrasound methods to address the cost and access constraints of MRI for the assessment, diagnosis, and monitoring of MASLD.

DiagnosticsVol. 16(5)
Rocky Vista University (US), University of Michigan (US), Gift of Life Michigan (US)
University of Michigan
Openalex Percentile: Top 30%
Liver Disease Diagnosis and Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.