Failure analysis of retrieved stainless steel Exeter hip implants: A fractographic and corrosion perspective

• Fatigue was the dominating failure mechanism for the three cemented REX 734 stems. • Cracks consistently initiated at the anterolateral region near the mid-length of the stem. • Coarse Nb-enriched phases likely assisted crack initiation and propagation. • Insufficient cement fixation likely enabled fluid ingress and corrosion fatigue. • Patient weight and BMI influenced crack propagation and the area of final fracture. In this study, three fractured stainless steel Exeter (REX 734) hip joint implants were analyzed to identify the underlying failure mechanisms. These cemented hip stems with a polished surface finish failed earlier than their expected service life, indicated by sudden pain in the patients, necessitating revision surgeries. Optical and scanning electron microscopy were used to analyze the explanted hip stem fracture surfaces. The fractograms exhibited three distinctive zones; (1) crack initiation at the anterolateral part of the stem, (2) crack propagation zone with beach marks and striations, and (3) final fracture zone. Fatigue was the dominant failure mechanism. The anterolateral surface of the implants close to the fracture initiation site showed discoloration and surface alteration, possibly due to corrosion, and the presence of extrusions and intrusions formed by slip band accumulation, causing stress concentration and crack initiation. Also, elongated and sharp-edged phases enriched in niobium were found at the crack initiation site and throughout the microstructure. Medical radiographs indicated a lack of cement fixation at the proximal part of the stems, causing implant instability and ingress of body fluids, which potentially led to corrosion reactions. The crack propagation zone decreased in relative size with higher patient weight and BMI. Overall, this study suggests the possibility of corrosion contributions and a role of niobium-rich precipitates in the crack initiation and fatigue failure.

Authors

Institutions

Publication Details

Journal
Engineering Failure Analysis
Published
2026-03-07
DOI
https://doi.org/10.1016/j.engfailanal.2026.110718
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Failure analysis of retrieved stainless steel Exeter hip implants: A fractographic and corrosion perspective

Saman Nikpour, Yolanda Susanne Hedberg, Matthew G. Teeter, Fuad Khoshnaw et al.
Engineering Failure Analysis
Orthopaedic implants and arthroplasty
article

Failure analysis of retrieved stainless steel Exeter hip implants: A fractographic and corrosion perspective

Saman Nikpour, Yolanda Susanne Hedberg, Matthew G. Teeter, Fuad Khoshnaw, Vahid Dehnavi, Brent Lanting
article en

Abstract

• Fatigue was the dominating failure mechanism for the three cemented REX 734 stems. • Cracks consistently initiated at the anterolateral region near the mid-length of the stem. • Coarse Nb-enriched phases likely assisted crack initiation and propagation. • Insufficient cement fixation likely enabled fluid ingress and corrosion fatigue. • Patient weight and BMI influenced crack propagation and the area of final fracture. In this study, three fractured stainless steel Exeter (REX 734) hip joint implants were analyzed to identify the underlying failure mechanisms. These cemented hip stems with a polished surface finish failed earlier than their expected service life, indicated by sudden pain in the patients, necessitating revision surgeries. Optical and scanning electron microscopy were used to analyze the explanted hip stem fracture surfaces. The fractograms exhibited three distinctive zones; (1) crack initiation at the anterolateral part of the stem, (2) crack propagation zone with beach marks and striations, and (3) final fracture zone. Fatigue was the dominant failure mechanism. The anterolateral surface of the implants close to the fracture initiation site showed discoloration and surface alteration, possibly due to corrosion, and the presence of extrusions and intrusions formed by slip band accumulation, causing stress concentration and crack initiation. Also, elongated and sharp-edged phases enriched in niobium were found at the crack initiation site and throughout the microstructure. Medical radiographs indicated a lack of cement fixation at the proximal part of the stems, causing implant instability and ingress of body fluids, which potentially led to corrosion reactions. The crack propagation zone decreased in relative size with higher patient weight and BMI. Overall, this study suggests the possibility of corrosion contributions and a role of niobium-rich precipitates in the crack initiation and fatigue failure.

Engineering Failure AnalysisVol. 190
Western University (CA), London Health Sciences Centre (CA), De Montfort University (GB)
Openalex Percentile: Top 30%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.