3D-DXA cortical and trabecular parameters: agreement and precision between GE Healthcare Prodigy and iDXA densitometers
Summary We assessed agreement and precision of 3D-DXA parameters between Prodigy and iDXA densitometers in adults. Strong agreement was observed between devices, and short-term precision was comparable. Moreover, high inter-scanner diagnostic classification agreement was found in the cohort spanning the full range of bone density categories, supporting the use of 3D-DXA across these GE Healthcare models for cross-sectional diagnostic purposes. Purpose 3D-DXA reconstructs DXA hip scans to 3-dimensional images, allowing measurement of trabecular and cortical bone parameters. Differences in imaging characteristics between the GE Healthcare iDXA and Prodigy could influence 3D-DXA results. This study evaluated the agreement and precision of 3D-DXA cortical and trabecular femur parameters between Prodigy and iDXA densitometers. Methods A total of 391 adults from 3 clinical centers in the United States and Brazil underwent hip scans on both Prodigy and iDXA scanners, subsequently analyzed with 3D-Shaper software. Short-term precision was assessed on three Prodigy and two iDXA densitometers. Agreement was assessed by regression, Bland-Altman analyses, and diagnostic classification agreement analysis. Short-term precision was determined following International Society for Clinical Densitometry recommendations. Results Bland-Altman analyses demonstrated statistically significant differences in aBMD and 3D-DXA measurements between Prodigy and iDXA scanners; however, diagnostic classification agreement was high. In the largest cohort, the only one spanning all three classification categories, agreement was 95.5% for cortical sBMD and 95.2% for trabecular vBMD (κw = 0.90 for both), comparable to total hip aBMD (97.3%, κw = 0.94). Short-term precision errors (RMS-SD) were 0.006 to 0.011 g/cm² for total hip aBMD, 0.009 to 0.023 g/cm² for femoral neck aBMD, 2.989 to 4.139 mg/cm³ for integral vBMD, 2.116 to 2.392 mg/cm² for cortical sBMD and 3.073 to 3.909 mg/cm³ for trabecular vBMD. Conclusion Under the conditions evaluated, aBMD and 3D-DXA bone measurements demonstrate comparable short-term precision and inter-scanner diagnostic classification agreement across Prodigy and iDXA densitometers.
Authors
- Miguel Madeira (ORCID: https://orcid.org/0000-0001-6752-2880)
- Marta Bracco
- Zhaojing Chen
- Neil Binkley (ORCID: https://orcid.org/0000-0002-9905-461X)
- Ludovic Humbert (ORCID: https://orcid.org/0000-0002-3675-7908)
- Luis del Río
- Silvana Di Gregorio (ORCID: https://orcid.org/0000-0001-6609-4335)
- Diane C. Krueger
Institutions
- Universidade Federal do Rio de Janeiro (BR)
- University of Wisconsin–Madison (US)
- Hospital Geral de Bonsucesso (BR)
- Hospital Universitário Clementino Fraga Filho (BR)
- California State University, San Bernardino (US)
Publication Details
- Journal
- Archives of Osteoporosis
- Published
- 2026-11-23
- DOI
- https://doi.org/10.1007/s11657-026-01811-9
- Citations
- 6
- Primary Topic
- Bone health and osteoporosis research
- Type
- article
- Field-Weighted Citation Impact
- 22.29