3D-DXA cortical and trabecular parameters: agreement and precision between GE Healthcare Prodigy and iDXA densitometers

Summary We assessed agreement and precision of 3D-DXA parameters between Prodigy and iDXA densitometers in adults. Strong agreement was observed between devices, and short-term precision was comparable. Moreover, high inter-scanner diagnostic classification agreement was found in the cohort spanning the full range of bone density categories, supporting the use of 3D-DXA across these GE Healthcare models for cross-sectional diagnostic purposes. Purpose 3D-DXA reconstructs DXA hip scans to 3-dimensional images, allowing measurement of trabecular and cortical bone parameters. Differences in imaging characteristics between the GE Healthcare iDXA and Prodigy could influence 3D-DXA results. This study evaluated the agreement and precision of 3D-DXA cortical and trabecular femur parameters between Prodigy and iDXA densitometers. Methods A total of 391 adults from 3 clinical centers in the United States and Brazil underwent hip scans on both Prodigy and iDXA scanners, subsequently analyzed with 3D-Shaper software. Short-term precision was assessed on three Prodigy and two iDXA densitometers. Agreement was assessed by regression, Bland-Altman analyses, and diagnostic classification agreement analysis. Short-term precision was determined following International Society for Clinical Densitometry recommendations. Results Bland-Altman analyses demonstrated statistically significant differences in aBMD and 3D-DXA measurements between Prodigy and iDXA scanners; however, diagnostic classification agreement was high. In the largest cohort, the only one spanning all three classification categories, agreement was 95.5% for cortical sBMD and 95.2% for trabecular vBMD (κw = 0.90 for both), comparable to total hip aBMD (97.3%, κw = 0.94). Short-term precision errors (RMS-SD) were 0.006 to 0.011 g/cm² for total hip aBMD, 0.009 to 0.023 g/cm² for femoral neck aBMD, 2.989 to 4.139 mg/cm³ for integral vBMD, 2.116 to 2.392 mg/cm² for cortical sBMD and 3.073 to 3.909 mg/cm³ for trabecular vBMD. Conclusion Under the conditions evaluated, aBMD and 3D-DXA bone measurements demonstrate comparable short-term precision and inter-scanner diagnostic classification agreement across Prodigy and iDXA densitometers.

Authors

Institutions

Publication Details

Journal
Archives of Osteoporosis
Published
2026-11-23
DOI
https://doi.org/10.1007/s11657-026-01811-9
Citations
6
Primary Topic
Bone health and osteoporosis research
Type
article
Field-Weighted Citation Impact
22.29
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
NOV
article

3D-DXA cortical and trabecular parameters: agreement and precision between GE Healthcare Prodigy and iDXA densitometers

Miguel Madeira, Marta Bracco, Zhaojing Chen, Neil Binkley et al.
6 citations
Archives of Osteoporosis
Bone health and osteoporosis research
22.29
article

3D-DXA cortical and trabecular parameters: agreement and precision between GE Healthcare Prodigy and iDXA densitometers

Miguel Madeira, Marta Bracco, Zhaojing Chen, Neil Binkley, Ludovic Humbert, Luis del Río, Silvana Di Gregorio, Diane C. Krueger
article en
6 citations

Abstract

Summary We assessed agreement and precision of 3D-DXA parameters between Prodigy and iDXA densitometers in adults. Strong agreement was observed between devices, and short-term precision was comparable. Moreover, high inter-scanner diagnostic classification agreement was found in the cohort spanning the full range of bone density categories, supporting the use of 3D-DXA across these GE Healthcare models for cross-sectional diagnostic purposes. Purpose 3D-DXA reconstructs DXA hip scans to 3-dimensional images, allowing measurement of trabecular and cortical bone parameters. Differences in imaging characteristics between the GE Healthcare iDXA and Prodigy could influence 3D-DXA results. This study evaluated the agreement and precision of 3D-DXA cortical and trabecular femur parameters between Prodigy and iDXA densitometers. Methods A total of 391 adults from 3 clinical centers in the United States and Brazil underwent hip scans on both Prodigy and iDXA scanners, subsequently analyzed with 3D-Shaper software. Short-term precision was assessed on three Prodigy and two iDXA densitometers. Agreement was assessed by regression, Bland-Altman analyses, and diagnostic classification agreement analysis. Short-term precision was determined following International Society for Clinical Densitometry recommendations. Results Bland-Altman analyses demonstrated statistically significant differences in aBMD and 3D-DXA measurements between Prodigy and iDXA scanners; however, diagnostic classification agreement was high. In the largest cohort, the only one spanning all three classification categories, agreement was 95.5% for cortical sBMD and 95.2% for trabecular vBMD (κw = 0.90 for both), comparable to total hip aBMD (97.3%, κw = 0.94). Short-term precision errors (RMS-SD) were 0.006 to 0.011 g/cm² for total hip aBMD, 0.009 to 0.023 g/cm² for femoral neck aBMD, 2.989 to 4.139 mg/cm³ for integral vBMD, 2.116 to 2.392 mg/cm² for cortical sBMD and 3.073 to 3.909 mg/cm³ for trabecular vBMD. Conclusion Under the conditions evaluated, aBMD and 3D-DXA bone measurements demonstrate comparable short-term precision and inter-scanner diagnostic classification agreement across Prodigy and iDXA densitometers.

Archives of OsteoporosisVol. 21(1)
Universidade Federal do Rio de Janeiro (BR), University of Wisconsin–Madison (US), Hospital Geral de Bonsucesso (BR), Hospital Universitário Clementino Fraga Filho (BR), California State University, San Bernardino (US)
Good health and well-being
Openalex Percentile: Top 0%
Bone health and osteoporosis research
22.29
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Laika access coming soon