Utilizing virtual reality to improve ICU visual displays
The ring-based display yielded improved alarm identification accuracy and response time compared to the graph-based alarm display. VR provided a robust, controlled method for testing medical interfaces. Future work should implement eye-tracking into alarm evaluation. This work's end goal is to combine our visual alarm with auditory alarms to create a multi-sensory alarm system.
Authors
- Joshua Shive (ORCID: https://orcid.org/0000-0002-5668-2149)
- Joseph J. Schlesinger (ORCID: https://orcid.org/0000-0003-3978-2647)
- Anqy Tong
- Hari Patel
- Sungmin Kwon
- Shayaan Essani
- Nathan Zhang
- Joy Zou
Institutions
- Vanderbilt University (US)
- Tennessee State University (US)
- Vanderbilt University Medical Center (US)
Publication Details
- Journal
- International Journal of Medical Informatics
- Published
- 2026-03-05
- DOI
- https://doi.org/10.1016/j.ijmedinf.2026.106366
- Primary Topic
- Healthcare Technology and Patient Monitoring
- Type
- article
- Field-Weighted Citation Impact