Estimation of the concentrations of operation room air pollutants and study of the various factors influencing their levels

Abstract Background and Aims: The air pollutants released in the operating room (OR) worsen the quality of air that the OR staff inhale and also exacerbate several illnesses among them. Material and Methods: Various OR pollutants were measured using portable air pollution monitoring equipment, including carbon dioxide (CO 2 ), total volatile organic compounds (TVOC), formaldehyde (HCHO), and particulate matter (PM2.5 and PM10). Results: The levels of PM2.5 (19.78 ± 8.81 mcg/m 3 vs. 29.83 ± 4.56 mcg/m 3 , P = 0.026) and PM10 (29.93 ± 12.96 mcg/m 3 vs. 38.61 ± 6.89 mcg/m 3 , P = 0.04) were significantly lower in the ORs with a high-efficiency particulate air (HEPA) filter compared to those without. The levels of other OR pollutants were comparable between the two groups. However, when cautery was used, the Avg TVOC (0.72 ± 0.3 mg/m 3 vs. 0.3 ± 0.28 mg/m 3 , P = 0.044), Avg PM2.5 (29.83 ± 4.57 mcg/m 3 vs. 19.78 ± 8 mcg/m 3 , P = 0.02), and Avg PM10 (38.61 ± 6.89 mcg/m 3 vs. 29.93 ± 12.96 mcg/m 3 , P = 0.044) were higher in the ORs compared to control. In ORs exposed to glutaraldehyde solution, the average TVOC (0.36 ± 0.32 vs. 0.72 ± 0.3, P = 0.04) and PM2.5 (29.83 ± 4.56 mcg/m 3 vs. 20.61 ± 9.70 mcg/m 3 , P = 0.04) were lower in the ORs compared to those of the control. The PM2.5 [36.66 (34.93,38.38)] μg/m 3 and PM10 [49.33 (46.88,51.78)] μg/m 3 levels of the control area are highest in winters but significantly lower in the ORs [PM2.5–17.0 (15.52, 18.47)] μg/m 3 and PM10 [25.02 (20.02,30.02)] μg/m 3 . Conclusions: The average CO 2 level in the ORs was often severe enough to cause cognitive dysfunction among healthcare workers (HCWs). The TVOC level at the anesthesia workstation was higher than the baseline. The levels of PM2.5 exceeded the annual average. The study highlights the need to improve air quality in ORs.

Authors

Institutions

Publication Details

Journal
Journal of Anaesthesiology Clinical Pharmacology
Published
2026-03-03
DOI
https://doi.org/10.4103/joacp.joacp_380_25
Primary Topic
Infection Control and Ventilation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Estimation of the concentrations of operation room air pollutants and study of the various factors influencing their levels

Rajani Rai, Mamta Harjai, SURAJ KUMAR, Soumya Sankar Nath
Journal of Anaesthesiology Clinical Pharmacology
Infection Control and Ventilation
article

Estimation of the concentrations of operation room air pollutants and study of the various factors influencing their levels

Rajani Rai, Mamta Harjai, SURAJ KUMAR, Soumya Sankar Nath
article en

Abstract

Abstract Background and Aims: The air pollutants released in the operating room (OR) worsen the quality of air that the OR staff inhale and also exacerbate several illnesses among them. Material and Methods: Various OR pollutants were measured using portable air pollution monitoring equipment, including carbon dioxide (CO 2 ), total volatile organic compounds (TVOC), formaldehyde (HCHO), and particulate matter (PM2.5 and PM10). Results: The levels of PM2.5 (19.78 ± 8.81 mcg/m 3 vs. 29.83 ± 4.56 mcg/m 3 , P = 0.026) and PM10 (29.93 ± 12.96 mcg/m 3 vs. 38.61 ± 6.89 mcg/m 3 , P = 0.04) were significantly lower in the ORs with a high-efficiency particulate air (HEPA) filter compared to those without. The levels of other OR pollutants were comparable between the two groups. However, when cautery was used, the Avg TVOC (0.72 ± 0.3 mg/m 3 vs. 0.3 ± 0.28 mg/m 3 , P = 0.044), Avg PM2.5 (29.83 ± 4.57 mcg/m 3 vs. 19.78 ± 8 mcg/m 3 , P = 0.02), and Avg PM10 (38.61 ± 6.89 mcg/m 3 vs. 29.93 ± 12.96 mcg/m 3 , P = 0.044) were higher in the ORs compared to control. In ORs exposed to glutaraldehyde solution, the average TVOC (0.36 ± 0.32 vs. 0.72 ± 0.3, P = 0.04) and PM2.5 (29.83 ± 4.56 mcg/m 3 vs. 20.61 ± 9.70 mcg/m 3 , P = 0.04) were lower in the ORs compared to those of the control. The PM2.5 [36.66 (34.93,38.38)] μg/m 3 and PM10 [49.33 (46.88,51.78)] μg/m 3 levels of the control area are highest in winters but significantly lower in the ORs [PM2.5–17.0 (15.52, 18.47)] μg/m 3 and PM10 [25.02 (20.02,30.02)] μg/m 3 . Conclusions: The average CO 2 level in the ORs was often severe enough to cause cognitive dysfunction among healthcare workers (HCWs). The TVOC level at the anesthesia workstation was higher than the baseline. The levels of PM2.5 exceeded the annual average. The study highlights the need to improve air quality in ORs.

Journal of Anaesthesiology Clinical Pharmacology
Dr. Ram Manohar Lohia Institute of Medical Sciences (IN)
Openalex Percentile: Top 35%
Infection Control and Ventilation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.