Systemic Cobalt Toxicity Secondary to Tibial Component Malpositioning in Bilateral Primary Total Knee Arthroplasties

Metallosis and systemic metal toxicity are uncommon complications in modern total joint arthroplasty, typically linked to metal-on-metal hip articulations. We present a case of systemic cobalt toxicity following bilateral total knee arthroplasties (TKAs), where excessive posterior tibial slope caused flexion gap instability and polyethylene wear with subsequent metallic debris release. This patient developed cognitive impairment, neuropathy, and cardiac symptoms nearly a decade after surgery, with the highest serum cobalt and chromium levels reported in literature for primary TKAs. At revision, polyethylene and baseplate wear with metallic debris and bone loss were present. Surgeons should remain vigilant for metallosis and systemic toxicity in painful or unstable TKAs with polyethylene wear from component malalignment or instability. Further studies are needed to define surveillance strategies and management guidelines.

Authors

Institutions

Publication Details

Journal
Arthroplasty Today
Published
2026-03-02
DOI
https://doi.org/10.1016/j.artd.2026.101973
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Systemic Cobalt Toxicity Secondary to Tibial Component Malpositioning in Bilateral Primary Total Knee Arthroplasties

Michael J. Archibeck, Nikhil Vallabhaneni, Kayla Gates, Jeremy M. Gililland et al.
Arthroplasty Today
Orthopaedic implants and arthroplasty
article

Systemic Cobalt Toxicity Secondary to Tibial Component Malpositioning in Bilateral Primary Total Knee Arthroplasties

Michael J. Archibeck, Nikhil Vallabhaneni, Kayla Gates, Jeremy M. Gililland, Joshua P. Rainey, Lucas A. Anderson
article en

Abstract

Metallosis and systemic metal toxicity are uncommon complications in modern total joint arthroplasty, typically linked to metal-on-metal hip articulations. We present a case of systemic cobalt toxicity following bilateral total knee arthroplasties (TKAs), where excessive posterior tibial slope caused flexion gap instability and polyethylene wear with subsequent metallic debris release. This patient developed cognitive impairment, neuropathy, and cardiac symptoms nearly a decade after surgery, with the highest serum cobalt and chromium levels reported in literature for primary TKAs. At revision, polyethylene and baseplate wear with metallic debris and bone loss were present. Surgeons should remain vigilant for metallosis and systemic toxicity in painful or unstable TKAs with polyethylene wear from component malalignment or instability. Further studies are needed to define surveillance strategies and management guidelines.

Arthroplasty TodayVol. 38
University of Utah (US)
Stryker, American Association of Hip and Knee Surgeons
Zero hunger
Openalex Percentile: Top 36%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Systemic Cobalt Toxicity Secondary to Tibial Component Malpositioning in Bilateral Primary Total Knee Arthroplasties — Michael J. Archibeck, Nikhil Vallabhaneni, et al. · Arthroplasty Today (2026) | TGRS Research Map | TGRS