Evaluation of peripheral intravenous line securement devices under clinically relevant loading and perspiratory conditions

Peripheral vascular access devices (PVADs) and their associated intravenous (IV) lines are widely used in clinical care. Securement devices are typically employed to prevent accidental dislodgement and complications such as occlusion and phlebitis. Despite the wide use of securement devices, there are no standardised methods for evaluating their actual ability to secure IV lines. This study assessed the securing ability of four commercially available IV line securement devices - Grip Lok, Javelo, micropore tape, and IV bracelet - under controlled laboratory conditions. First, all four devices were tested for resistance to axial pull-out forces using a uniaxial materials testing machine across ten participants. Participant comfort during testing was assessed via questionnaire. Grip-Lok and Javelo exhibited significantly greater axial pull-out strength than the other two devices (p-adj < 0.001), and were progressed to further testing, consisting of off-axis loading at 90°, 135°, and peel-off, and axial pull-out under simulated perspiration conditions. No significant differences were observed between Grip Lok and Javelo under loading in 90° and 135°. Javelo demonstrated superior performance in both peel-off and perspirant conditions (p-adj < 0.05). Javelo was rated the most comfortable by 60% of participants. The results highlight the need for incorporating realistic loading conditions and comfort metrics when evaluating the performance of infusion line securement aids. Furthermore, the results provide evidence that using a non-adhesive based device, particularly in patients with marked diaphoresis might provide more infusion line security, though further study is required.

Authors

Institutions

Publication Details

Journal
PLoS ONE
Published
2026-03-02
DOI
https://doi.org/10.1371/journal.pone.0341618
Primary Topic
Central Venous Catheters and Hemodialysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of peripheral intravenous line securement devices under clinically relevant loading and perspiratory conditions

Maegan Spiteri, George K. H. Morgan, Spyros Masouros
PLoS ONE
Central Venous Catheters and Hemodialysis
article

Evaluation of peripheral intravenous line securement devices under clinically relevant loading and perspiratory conditions

Maegan Spiteri, George K. H. Morgan, Spyros Masouros
article en

Abstract

Peripheral vascular access devices (PVADs) and their associated intravenous (IV) lines are widely used in clinical care. Securement devices are typically employed to prevent accidental dislodgement and complications such as occlusion and phlebitis. Despite the wide use of securement devices, there are no standardised methods for evaluating their actual ability to secure IV lines. This study assessed the securing ability of four commercially available IV line securement devices - Grip Lok, Javelo, micropore tape, and IV bracelet - under controlled laboratory conditions. First, all four devices were tested for resistance to axial pull-out forces using a uniaxial materials testing machine across ten participants. Participant comfort during testing was assessed via questionnaire. Grip-Lok and Javelo exhibited significantly greater axial pull-out strength than the other two devices (p-adj < 0.001), and were progressed to further testing, consisting of off-axis loading at 90°, 135°, and peel-off, and axial pull-out under simulated perspiration conditions. No significant differences were observed between Grip Lok and Javelo under loading in 90° and 135°. Javelo demonstrated superior performance in both peel-off and perspirant conditions (p-adj < 0.05). Javelo was rated the most comfortable by 60% of participants. The results highlight the need for incorporating realistic loading conditions and comfort metrics when evaluating the performance of infusion line securement aids. Furthermore, the results provide evidence that using a non-adhesive based device, particularly in patients with marked diaphoresis might provide more infusion line security, though further study is required.

PLoS ONEVol. 21(3)
Imperial College London (GB)
Good health and well-being
Openalex Percentile: Top 44%
Central Venous Catheters and Hemodialysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.