A dual-function dry electrode for electromyography recording and transcutaneous electrical stimulation

Electromyography and transcutaneous electrical stimulation are widely used for neuromuscular assessment and therapy, respectively. However, when the target muscle group is small, such as in post-stroke rehabilitation, there is limited space to place separate electrodes for sensing and stimulation. Developing an integrated electrode capable of both functions would address this limitation by enabling precise monitoring and targeted stimulation within the same localized region, while also simplifying system design and improving functional outcomes. Additionally, most current electrodes are wet and require gel, which can be inconvenient for users. This study explored the feasibility of using a dual-function dry electrode to both record electromyography signals and deliver transcutaneous electrical stimulation. To evaluate the practical effectiveness of the electrode for functional applications, a hand gesture recognition algorithm was developed to assess the quality and reliability of the electromyography signals recorded by the electrode. Forty participants wore an armband equipped with these electrodes and performed six different hand gestures. The algorithm achieved a recognition accuracy of 80.5%, which was comparable to the accuracy of the commercial Myo armband-based hand gesture recognition algorithm. The performance of the dual-function electrode in delivering electrical stimulation was assessed by measuring stimulation intensity at three threshold levels across all 40 participants and by analyzing the induced torque data due to stimulation. Results indicated no significant difference in stimulation intensity and the induced torque measurements when compared to those obtained with a commercial electrode ([Formula: see text]). These findings show potential in integrating transcutaneous electrical stimulation and electromyography functionalities into a single dry electrode, offering a path toward enhanced user comfort, reduced costs, and improved therapeutic interventions.

Authors

Institutions

Publication Details

Journal
Scientific Reports
Published
2026-03-02
DOI
https://doi.org/10.1038/s41598-026-41729-3
Primary Topic
Muscle activation and electromyography studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A dual-function dry electrode for electromyography recording and transcutaneous electrical stimulation

Maedeh Mohammadiazni, Ana Luisa Trejos, Yue Zhou
Scientific Reports
Muscle activation and electromyography studies
article

A dual-function dry electrode for electromyography recording and transcutaneous electrical stimulation

Maedeh Mohammadiazni, Ana Luisa Trejos, Yue Zhou
article en

Abstract

Electromyography and transcutaneous electrical stimulation are widely used for neuromuscular assessment and therapy, respectively. However, when the target muscle group is small, such as in post-stroke rehabilitation, there is limited space to place separate electrodes for sensing and stimulation. Developing an integrated electrode capable of both functions would address this limitation by enabling precise monitoring and targeted stimulation within the same localized region, while also simplifying system design and improving functional outcomes. Additionally, most current electrodes are wet and require gel, which can be inconvenient for users. This study explored the feasibility of using a dual-function dry electrode to both record electromyography signals and deliver transcutaneous electrical stimulation. To evaluate the practical effectiveness of the electrode for functional applications, a hand gesture recognition algorithm was developed to assess the quality and reliability of the electromyography signals recorded by the electrode. Forty participants wore an armband equipped with these electrodes and performed six different hand gestures. The algorithm achieved a recognition accuracy of 80.5%, which was comparable to the accuracy of the commercial Myo armband-based hand gesture recognition algorithm. The performance of the dual-function electrode in delivering electrical stimulation was assessed by measuring stimulation intensity at three threshold levels across all 40 participants and by analyzing the induced torque data due to stimulation. Results indicated no significant difference in stimulation intensity and the induced torque measurements when compared to those obtained with a commercial electrode ([Formula: see text]). These findings show potential in integrating transcutaneous electrical stimulation and electromyography functionalities into a single dry electrode, offering a path toward enhanced user comfort, reduced costs, and improved therapeutic interventions.

Scientific Reports
Western University (CA)
Affordable and clean energy
Openalex Percentile: Top 39%
Muscle activation and electromyography studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

A dual-function dry electrode for electromyography recording and transcutaneous electrical stimulation — Maedeh Mohammadiazni, Ana Luisa Trejos, et al. · Scientific Reports (2026) | TGRS Research Map | TGRS