Many bioanalytical methods result in the generation of colorimetric readouts, with color intensity providing information of the concentration of the targeted analyte. However, conventional absorbance readers often saturate at high analyte concentrations, requiring sample serial dilutions when the analyte studied can be present in a wide range of concentrations. Here, we present a portable reader based on a single-photon avalanche diode (SPAD) array performing in-pixel time-resolved photon counting to quantify absorbance over an extended dynamic range. The method relies on a lifetime-based measurement principle, in which absorbance is derived from the photon-arrival decay constant (τ) rather than steady-state intensity, enabling accurate quantification under high-flux conditions. We applied this approach to an influenza B nucleoprotein magnetic particle immunoassay. The SPAD-based reader maintained a monotonic response up to 64 ng mL<sup>-1</sup> within the useable dynamic range, compared to signal saturation above 12.5 ng mL<sup>-1</sup> for a commercial photometer, with limits of detection (LoD) and quantification (LoQ) of 0.0034 and 0.0949 ng mL<sup>-1</sup>, respectively, while maintaining coefficients of variation (CV) < 10%. A convolutional neural network trained on 16-bin histograms accurately inferred analyte concentration (R<sup>2</sup> = 0.986), yielding results consistent with the τ-based calibration, thus demonstrating that the same time-resolved photon information can be quantitatively interpreted either through physical modeling or data-driven inference. By extending the useable absorbance range while preserving sensitivity, this approach improves key analytical figures of merit compared to commercial benchtop equipment, paving the way for high-performance portable point-of-care (PoC) solutions.

Authors

Institutions

Publication Details

Journal
Talanta
Published
2026-03-02
DOI
https://doi.org/10.1016/j.talanta.2026.129611
Primary Topic
Polydiacetylene-based materials and applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Portable colorimetric absorbance reader with extended dynamic range via time-resolved sensing

A. Diéguez, O. Caravaca-Muller, S. Moreno, M.B. Jaballah et al.
Talanta
Polydiacetylene-based materials and applications
article

Portable colorimetric absorbance reader with extended dynamic range via time-resolved sensing

A. Diéguez, O. Caravaca-Muller, S. Moreno, M.B. Jaballah, A. Vilà, E. Baldrich, X. García-Fernández
article en

Abstract

Many bioanalytical methods result in the generation of colorimetric readouts, with color intensity providing information of the concentration of the targeted analyte. However, conventional absorbance readers often saturate at high analyte concentrations, requiring sample serial dilutions when the analyte studied can be present in a wide range of concentrations. Here, we present a portable reader based on a single-photon avalanche diode (SPAD) array performing in-pixel time-resolved photon counting to quantify absorbance over an extended dynamic range. The method relies on a lifetime-based measurement principle, in which absorbance is derived from the photon-arrival decay constant (τ) rather than steady-state intensity, enabling accurate quantification under high-flux conditions. We applied this approach to an influenza B nucleoprotein magnetic particle immunoassay. The SPAD-based reader maintained a monotonic response up to 64 ng mL<sup>-1</sup> within the useable dynamic range, compared to signal saturation above 12.5 ng mL<sup>-1</sup> for a commercial photometer, with limits of detection (LoD) and quantification (LoQ) of 0.0034 and 0.0949 ng mL<sup>-1</sup>, respectively, while maintaining coefficients of variation (CV) < 10%. A convolutional neural network trained on 16-bin histograms accurately inferred analyte concentration (R<sup>2</sup> = 0.986), yielding results consistent with the τ-based calibration, thus demonstrating that the same time-resolved photon information can be quantitatively interpreted either through physical modeling or data-driven inference. By extending the useable absorbance range while preserving sensitivity, this approach improves key analytical figures of merit compared to commercial benchtop equipment, paving the way for high-performance portable point-of-care (PoC) solutions.

TalantaVol. 305
Vall d'Hebron Institut de Recerca (ES), Centro de Investigación Biomédica en Red (ES), Institut de Nanociència i Nanotecnologia de la Universitat de Barcelona, Universitat de Barcelona (ES)
Affordable and clean energy
Openalex Percentile: Top 37%
Polydiacetylene-based materials and applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.