Diagnostic performance evaluation of the Siemens Healthineers Atellica HEMA 580 hematology analyzer for abnormal cell detection and optimization of immature myeloid cell cut-offs
The Atellica HEMA 580 provided reliable performance in detecting immature myeloid cells and NRBCs. The IMG cut-off >0.8, determined through ROC optimization, supports its implementation for enhancing laboratory workflow efficiency while maintaining diagnostic accuracy. These findings support the clinical implementation of the Atellica HEMA 580 for routine hematology laboratories, particularly for enhanced detection of clinically significant immature myeloid populations.
Authors
- Vincenzo De Iuliis (ORCID: https://orcid.org/0000-0001-7479-7452)
Institutions
- Nini Hospital (LB)
Publication Details
- Journal
- Clinical Chemistry and Laboratory Medicine (CCLM)
- Published
- 2026-03-02
- DOI
- https://doi.org/10.1515/cclm-2026-0083
- Primary Topic
- Clinical Laboratory Practices and Quality Control
- Type
- article
- Field-Weighted Citation Impact
- 0.00