Development and Validation of a Wireless, Low-Cost Device for Dual Measurement of in-Shoe Plantar Pressure and Temperature in High-Risk Diabetic Feet
This study aims to evaluate the performance of an innovative in-shoe sensing device against the FScan system. Across five trials, the novel device demonstrated consistently strong correlations with FScan peak pressure readings (<i>r</i> = 0.801, <i>r</i> = 0.978, <i>r</i> = 0.813, <i>r</i> = 0.887, <i>r</i> = 0.944). Superimposed peak-pressure plots revealed highly similar waveform patterns between the two systems. Error-based metrics also supported this equivalence. Root Mean Squared Logarithmic Error values similarly suggested consistent predictive agreement. Temperature-sensing performance was assessed by comparing in-shoe sensor readings by the novel device with thermal-camera measurements. The thermal camera detected an average temperature change of 3.7°C, whereas the in-shoe sensor recorded an average change of 0.67°C. There was substantial variability in both systems, with higher variation observed in the in-shoe sensor than in the thermal camera. Despite variability, temperature and pressure measurements from the novel device were strongly correlated (<i>r</i> = 0.87). In conclusion, the new device provides pressure measurements comparable to the FScan system and appears promising for monitoring foot health.
Authors
- Cynthia Formosa (ORCID: https://orcid.org/0000-0002-4251-1621)
- Claire Saliba Thorne (ORCID: https://orcid.org/0000-0003-2202-7098)
- Alfred Gatt (ORCID: https://orcid.org/0000-0002-0489-6479)
- Geoffrey Attard
- Nikolaos Papanas
- Clifford DeRaffaele
Institutions
- Democritus University of Thrace (GR)
- University of Malta (MT)
Publication Details
- Journal
- The International Journal of Lower Extremity Wounds
- Published
- 2026-03-02
- DOI
- https://doi.org/10.1177/15347346261428561
- Primary Topic
- Diabetic Foot Ulcer Assessment and Management
- Type
- article
- Field-Weighted Citation Impact
- 0.00