Surgical Performance of 3D-Printed Polyetheretherketone (PEEK) Patient-Specific Implants and Titanium Mesh in Clinically Matched Orbital Reconstruction: A Cadaveric Study

Orbital reconstruction following trauma remains challenging due to complex three-dimensional (3D) anatomy and limited surgical access. While pre-fabricated titanium mesh is standard, it requires extensive intraoperative manipulation and produces imaging artifacts. The 3D-printed polyetheretherketone (PEEK) patient-specific implants (PSIs) offer potential advantages; however, limited data exists for the acceptance of PEEK PSIs by surgeons compared to other established techniques. Fourteen surgeons performed simulated orbital reconstructions on nine cadaveric heads comparing titanium mesh and the 3D-printed PEEK PSIs. Titanium mesh was used for Class II orbital floor fractures, while the 3D-printed PEEK PSIs (native and radiopaque formulations) were used for Class IV defects. Surgeons were blinded to the PEEK formulation type. Outcomes included operative efficiency, handling characteristics, fit quality, and mechanical stability using validated 5-point Likert scales and objective timing. The 3D-printed PEEK PSIs demonstrated faster procedure times (9.5 ± 5.3 vs. 11.2 ± 5.1 min) and superior fit quality (2.00 ± 1.04 vs. 2.18 ± 0.60) and mechanical stability (1.67 ± 0.49 vs. 1.91 ± 0.54), with 100% rated stable versus 91% for the titanium mesh. Surgeons could not distinguish between the native and radiopaque PEEK formulations. Most surgeons (64.3%) preferred situation-dependent material selection. The 3D-printed PEEK PSIs demonstrated advantages in handling, fit quality, and mechanical stability for complex defects, while the titanium mesh showed a lower learning curve for simple reconstructions. Radiopaque enhancement expands PEEK's clinical utility without compromising handling.

Authors

Institutions

Publication Details

Journal
Craniomaxillofacial Trauma & Reconstruction
Published
2026-03-02
DOI
https://doi.org/10.3390/cmtr19010013
Primary Topic
Facial Trauma and Fracture Management
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Surgical Performance of 3D-Printed Polyetheretherketone (PEEK) Patient-Specific Implants and Titanium Mesh in Clinically Matched Orbital Reconstruction: A Cadaveric Study

Neha Sharma, Florian M. Thieringer, Dominik Haenggi, Jokin Zubizarreta Oteiza et al.
Craniomaxillofacial Trauma & Reconstruction
Facial Trauma and Fracture Management
article

Surgical Performance of 3D-Printed Polyetheretherketone (PEEK) Patient-Specific Implants and Titanium Mesh in Clinically Matched Orbital Reconstruction: A Cadaveric Study

Neha Sharma, Florian M. Thieringer, Dominik Haenggi, Jokin Zubizarreta Oteiza, Daniel Seiler, Lukas Schuebel, Yannick Simon Krieger
article en

Abstract

Orbital reconstruction following trauma remains challenging due to complex three-dimensional (3D) anatomy and limited surgical access. While pre-fabricated titanium mesh is standard, it requires extensive intraoperative manipulation and produces imaging artifacts. The 3D-printed polyetheretherketone (PEEK) patient-specific implants (PSIs) offer potential advantages; however, limited data exists for the acceptance of PEEK PSIs by surgeons compared to other established techniques. Fourteen surgeons performed simulated orbital reconstructions on nine cadaveric heads comparing titanium mesh and the 3D-printed PEEK PSIs. Titanium mesh was used for Class II orbital floor fractures, while the 3D-printed PEEK PSIs (native and radiopaque formulations) were used for Class IV defects. Surgeons were blinded to the PEEK formulation type. Outcomes included operative efficiency, handling characteristics, fit quality, and mechanical stability using validated 5-point Likert scales and objective timing. The 3D-printed PEEK PSIs demonstrated faster procedure times (9.5 ± 5.3 vs. 11.2 ± 5.1 min) and superior fit quality (2.00 ± 1.04 vs. 2.18 ± 0.60) and mechanical stability (1.67 ± 0.49 vs. 1.91 ± 0.54), with 100% rated stable versus 91% for the titanium mesh. Surgeons could not distinguish between the native and radiopaque PEEK formulations. Most surgeons (64.3%) preferred situation-dependent material selection. The 3D-printed PEEK PSIs demonstrated advantages in handling, fit quality, and mechanical stability for complex defects, while the titanium mesh showed a lower learning curve for simple reconstructions. Radiopaque enhancement expands PEEK's clinical utility without compromising handling.

Craniomaxillofacial Trauma & ReconstructionVol. 19(1)
University of Basel (CH), University Hospital of Basel (CH), FHNW University of Applied Sciences and Arts Northwestern Switzerland (CH), SingHealth Duke-NUS Academic Medical Centre (SG), Dassault Systèmes (Germany) (DE)
International Bone Research Association
Openalex Percentile: Top 37%
Facial Trauma and Fracture Management
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.