Automated administration of medical oxygen using model predictive control incorporating real-time monitoring of breathing parameters

Current automated oxygen delivery systems for spontaneously breathing patients rely solely on peripheral oxygen saturation monitoring and do not account for the influence of breathing parameters on the fraction of inspired oxygen delivered via nasal cannula. This study incorporated breathing parameters that affect the fraction of inspired oxygen, namely, tidal volume and inhalation time, alongside peripheral oxygen saturation monitoring to design a closed-loop control system. A model of respiratory and circulatory systems presented in the literature was used to simulate the peripheral oxygen saturation under patient-specific pathophysiological conditions. A model predictive controller was used to incorporate both breathing parameters and peripheral oxygen saturation measurements into the control system design. For eight different cases, the efficiency of the model predictive controller was compared with that of a tuned proportional-integral-derivative controller, which relied solely on peripheral oxygen saturation monitoring. The model predictive controller achieved a higher efficiency score defined as the difference between time spent in the target range of the peripheral oxygen saturation (between 88% and 92%), and average nasal cannula flow (mL/s) (mean = 0.80 ± 0.07), compared to a proportional-integral-derivative controller (0.73 ± 0.08). The model predictive controller showed the best balance between accurate oxygenation and oxygen consumption with the lowest variability. These findings suggest that a model predictive controller integrating both breathing parameters and peripheral oxygen saturation monitoring offers a reliable and adaptive approach for home-based oxygen therapy in patients with chronic obstructive pulmonary disease, reducing the need for retuning and improving oxygen delivery by balancing oxygenation and oxygen use.

Authors

Institutions

Publication Details

Journal
Computers in Biology and Medicine
Published
2026-03-02
DOI
https://doi.org/10.1016/j.compbiomed.2026.111592
Primary Topic
Healthcare Technology and Patient Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated administration of medical oxygen using model predictive control incorporating real-time monitoring of breathing parameters

Hossein Rouhani, Jakub Ráfl, Mozhgan Sabz, Edmund Morozoff et al.
Computers in Biology and Medicine
Healthcare Technology and Patient Monitoring
article

Automated administration of medical oxygen using model predictive control incorporating real-time monitoring of breathing parameters

Hossein Rouhani, Jakub Ráfl, Mozhgan Sabz, Edmund Morozoff, Andrew Martin
article en

Abstract

Current automated oxygen delivery systems for spontaneously breathing patients rely solely on peripheral oxygen saturation monitoring and do not account for the influence of breathing parameters on the fraction of inspired oxygen delivered via nasal cannula. This study incorporated breathing parameters that affect the fraction of inspired oxygen, namely, tidal volume and inhalation time, alongside peripheral oxygen saturation monitoring to design a closed-loop control system. A model of respiratory and circulatory systems presented in the literature was used to simulate the peripheral oxygen saturation under patient-specific pathophysiological conditions. A model predictive controller was used to incorporate both breathing parameters and peripheral oxygen saturation measurements into the control system design. For eight different cases, the efficiency of the model predictive controller was compared with that of a tuned proportional-integral-derivative controller, which relied solely on peripheral oxygen saturation monitoring. The model predictive controller achieved a higher efficiency score defined as the difference between time spent in the target range of the peripheral oxygen saturation (between 88% and 92%), and average nasal cannula flow (mL/s) (mean = 0.80 ± 0.07), compared to a proportional-integral-derivative controller (0.73 ± 0.08). The model predictive controller showed the best balance between accurate oxygenation and oxygen consumption with the lowest variability. These findings suggest that a model predictive controller integrating both breathing parameters and peripheral oxygen saturation monitoring offers a reliable and adaptive approach for home-based oxygen therapy in patients with chronic obstructive pulmonary disease, reducing the need for retuning and improving oxygen delivery by balancing oxygenation and oxygen use.

Computers in Biology and MedicineVol. 205
University of Alberta (CA), Simon Fraser University (CA), Czech Technical University in Prague (CZ)
Zero hunger
Openalex Percentile: Top 37%
Healthcare Technology and Patient Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.