Enhancing pre-clinical microbeam radiation therapy capabilities at PETRA III beamline P61A through high-precision beam delivery

<i>Objective.</i>Microbeam radiation therapy (MRT) exploits spatially fractionated, micrometer-scale x-ray beam arrays that deliver high-dose (peak) and low-dose (valley) regions. At beamline P61A of the PETRA III synchrotron (DESY, Hamburg, Germany), MRT delivery has been limited to a fixed 2 mm wide field and single-direction irradiation. This study aimed to implement stereotactic multi-port MRT (STX-MRT) with lateral field extension and to quantify the spatial reproducibility of the microbeam arrays during patched and multi-angle delivery.<i>Approach.</i>A custom motorized stage providing vertical and lateral translation and rotation was developed to enable discrete lateral patching and sequential delivery from multiple ports (1-9). Each microbeam array comprised five peaks with a beam width of 50<i>µ</i>m and a center-to-center spacing of 400<i>µ</i>m. Spatial accuracy was assessed with Gafchromic™ HD-V2 film scanned at 7.9<i>µ</i>m resolution. Fifteen films (three per geometry), corresponding to 225 patched arrays, were included to determine the spatial accuracy.<i>Main results.</i>Beam delivery was highly reproducible. In 88% of the microbeam arrays, deviations were below the detection limit (⩽7.9<i>µ</i>m). Detectable displacements occurred in 12% of arrays and ranged from 33 to 258<i>µ</i>m, typically localized to individual patches or ports. The rotational isocenter remained stable across all configurations, and increasing the number of ports and delivery speed did not introduce systematic alignment errors.<i>Significance.</i>Reliable lateral field patching and multi-port MRT delivery with micrometer-scale precision was achieved, extending the usable irradiation field and enabling STX-MRT at beamline P61A. This technical capability broadens the range of irradiation of different target sizes and supports advanced preclinical studies of MRT.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2026-03-02
DOI
https://doi.org/10.1088/1361-6560/ae4c13
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Enhancing pre-clinical microbeam radiation therapy capabilities at PETRA III beamline P61A through high-precision beam delivery

Yvonne Dzierma, Akira Hamada, M. Kügele, Johannes Klingenberg et al.
Physics in Medicine and Biology
Advanced Radiotherapy Techniques
article

Enhancing pre-clinical microbeam radiation therapy capabilities at PETRA III beamline P61A through high-precision beam delivery

Yvonne Dzierma, Akira Hamada, M. Kügele, Johannes Klingenberg, V. T. de Rover, Bernd Frerker, Stefan Fiedler, Guido Hildebrandt, Elisabeth Schültke, Guilherme Abreu Faria
article en

Abstract

<i>Objective.</i>Microbeam radiation therapy (MRT) exploits spatially fractionated, micrometer-scale x-ray beam arrays that deliver high-dose (peak) and low-dose (valley) regions. At beamline P61A of the PETRA III synchrotron (DESY, Hamburg, Germany), MRT delivery has been limited to a fixed 2 mm wide field and single-direction irradiation. This study aimed to implement stereotactic multi-port MRT (STX-MRT) with lateral field extension and to quantify the spatial reproducibility of the microbeam arrays during patched and multi-angle delivery.<i>Approach.</i>A custom motorized stage providing vertical and lateral translation and rotation was developed to enable discrete lateral patching and sequential delivery from multiple ports (1-9). Each microbeam array comprised five peaks with a beam width of 50<i>µ</i>m and a center-to-center spacing of 400<i>µ</i>m. Spatial accuracy was assessed with Gafchromic™ HD-V2 film scanned at 7.9<i>µ</i>m resolution. Fifteen films (three per geometry), corresponding to 225 patched arrays, were included to determine the spatial accuracy.<i>Main results.</i>Beam delivery was highly reproducible. In 88% of the microbeam arrays, deviations were below the detection limit (⩽7.9<i>µ</i>m). Detectable displacements occurred in 12% of arrays and ranged from 33 to 258<i>µ</i>m, typically localized to individual patches or ports. The rotational isocenter remained stable across all configurations, and increasing the number of ports and delivery speed did not introduce systematic alignment errors.<i>Significance.</i>Reliable lateral field patching and multi-port MRT delivery with micrometer-scale precision was achieved, extending the usable irradiation field and enabling STX-MRT at beamline P61A. This technical capability broadens the range of irradiation of different target sizes and supports advanced preclinical studies of MRT.

Physics in Medicine and BiologyVol. 71(5)
University of Wollongong (AU), Skåne University Hospital (SE), Helmholtz-Zentrum Hereon (DE), European Molecular Biology Laboratory (DE), University of Rostock (DE)
Openalex Percentile: Top 33%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.