Background / Objective T1 mapping is widely used for tissue characterisation of the myocardium in cardiac MRI scans. Additionally, it can be a quantitative criterion in follow-up scans to assess interval change. High reproducibility of T1 mapping is crucial to ensure its usefulness in clinical practice as a follow-up assessment. However, the T1 mapping values depend on various factors (MRI field strength, scan sequence, heart rate, etc.) and significant artifacts may occur. The aim of this study was to assess the reproducibility of myocardial regional T1 mapping measurements in the interventricular septum using identical scan parameters and slice positions over a short time period, which on the 3T scanner is quite sparse elucidated. Methods Study cohort consisted of 51 patients who underwent cardiac MRI on 3T scanner. Native T1 mapping images were acquired in short-axis plane at three levels (basal, middle, and apical locations) in two time points: after scout sequences acquisition and at the end of the study or before intravenous contrast injection. The identical planes and saturation region for these measurements were applied. T1 mapping measurements were conducted by placing identical ROI’s of 1cm2 in the identical interventricular septum locations. Results The mean time interval between the two T1 mapping acquisitions was 19.4±5.6 minutes. No statistically significant difference between two T1 mapping values were observed. Reproducibility for interventricular septum was excellent for all slice position, with the highest one in the midventricular slice (intraclass correlation [ICC] 0.95), followed by basal slice (ICC 0.93)) and apical slice (ICC 0.92). F-test (0.7) did not find a statistically significant difference in the ICCs for the midventricular and apical slices. Conclusions T1 mapping values in a short time follow-up demonstrated excellent reproducibility in all locations of interventricular septum, although the highest one was observed in the midventricular slice and the lowest in the apical slice. Our findings are in line with results of Dekkers et. al. (J. MAGN. RESON. IMAGING 2019), which also found best reproducibility in the midventricular slice of septum, although our study demonstrated better intraclass correlation and lower difference of the reproducibility to the basal and apical septal slices.

Authors

Publication Details

Journal
Lithuanian University of Health Sciences
Published
2024-10-17
Primary Topic
Cardiac Imaging and Diagnostics
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reproducibility of T1 mapping measurements of interventricular septum in short term follow up on 3T MRI scanner

Valdonė Strelčiūnienė, Audronė Vaitiekienė, Lina Padervinskienė, Paulius Bučius et al.
Lithuanian University of Health Sciences
Cardiac Imaging and Diagnostics
article

Reproducibility of T1 mapping measurements of interventricular septum in short term follow up on 3T MRI scanner

Valdonė Strelčiūnienė, Audronė Vaitiekienė, Lina Padervinskienė, Paulius Bučius, Audra Banišauskaitė, Karolina Kozlovskytė, Neringa Gulbinaitė, Greta Jurkevičiūtė-Kučinskienė, Antanas Jankauskas
article en

Abstract

Background / Objective T1 mapping is widely used for tissue characterisation of the myocardium in cardiac MRI scans. Additionally, it can be a quantitative criterion in follow-up scans to assess interval change. High reproducibility of T1 mapping is crucial to ensure its usefulness in clinical practice as a follow-up assessment. However, the T1 mapping values depend on various factors (MRI field strength, scan sequence, heart rate, etc.) and significant artifacts may occur. The aim of this study was to assess the reproducibility of myocardial regional T1 mapping measurements in the interventricular septum using identical scan parameters and slice positions over a short time period, which on the 3T scanner is quite sparse elucidated. Methods Study cohort consisted of 51 patients who underwent cardiac MRI on 3T scanner. Native T1 mapping images were acquired in short-axis plane at three levels (basal, middle, and apical locations) in two time points: after scout sequences acquisition and at the end of the study or before intravenous contrast injection. The identical planes and saturation region for these measurements were applied. T1 mapping measurements were conducted by placing identical ROI’s of 1cm2 in the identical interventricular septum locations. Results The mean time interval between the two T1 mapping acquisitions was 19.4±5.6 minutes. No statistically significant difference between two T1 mapping values were observed. Reproducibility for interventricular septum was excellent for all slice position, with the highest one in the midventricular slice (intraclass correlation [ICC] 0.95), followed by basal slice (ICC 0.93)) and apical slice (ICC 0.92). F-test (0.7) did not find a statistically significant difference in the ICCs for the midventricular and apical slices. Conclusions T1 mapping values in a short time follow-up demonstrated excellent reproducibility in all locations of interventricular septum, although the highest one was observed in the midventricular slice and the lowest in the apical slice. Our findings are in line with results of Dekkers et. al. (J. MAGN. RESON. IMAGING 2019), which also found best reproducibility in the midventricular slice of septum, although our study demonstrated better intraclass correlation and lower difference of the reproducibility to the basal and apical septal slices.

Lithuanian University of Health Sciences
Openalex Percentile: Top 42%
Cardiac Imaging and Diagnostics
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.