<b>Background/Objectives:</b> Pediatric neurosurgical procedures often involve significant blood loss and rapid hemodynamic shifts, necessitating accurate hemoglobin (Hb) monitoring. While continuous non-invasive Hb (SpHb) monitoring offers real-time trending, its accuracy in high-risk pediatric populations remains debated. We aimed to evaluate the diagnostic accuracy and clinical utility of SpHb compared to invasive arterial blood gas (ABG) analysis in pediatric patients undergoing cranial and spinal surgeries. <b>Methods:</b> This prospective, observational study enrolled 60 pediatric patients (aged 0-16 years) scheduled for high-risk neurosurgery. SpHb was measured continuously and compared with intermittent ABG-Hb values. Statistical analysis included Bland-Altman agreement, Pearson's correlation, and Error Grid Analysis. Subgroup analyses assessed the impact of the Perfusion Index (PI), hypotension, and metabolic acidosis on device performance. <b>Results:</b> Data from 57 patients (median age: 12 months, interquartile range: 6-42 months; 70.2% aged <24 months) were analyzed. SpHb demonstrated a moderate correlation with ABG-Hb (r = 0.567, <i>p</i> < 0.001) but exhibited systematic overestimation with a mean bias of +1.60 ± 1.54 g/dL. Crucially, SpHb showed 0% sensitivity for detecting critical anemia (Hb < 8.0 g/dL). Device performance was significantly compromised by physiological extremes: severe metabolic acidosis significantly increased bias to +2.27 g/dL (<i>p</i> = 0.038), and intraoperative hypotension significantly widened the limits of agreement (SD of bias: 1.79 g/dL vs. 1.45 g/dL in normotension). Furthermore, hemodynamic analysis revealed a loss of autoregulation during hypotension, where the pressure-perfusion coupling strengthened (r = 0.44) compared to the normotensive state (r = 0.15). <b>Conclusions:</b> SpHb monitoring provides fair Hb trending but is limited by systematic overestimation and poor sensitivity for critical anemia. Accuracy worsens during severe acidosis and hemodynamic instability. Therefore, SpHb should function as a complementary "early warning" trend monitor rather than a sole transfusion trigger, with invasive validation remaining essential for intraoperative decision-making.

Authors

Institutions

Publication Details

Journal
Children
Published
2026-02-25
DOI
https://doi.org/10.3390/children13030323
Primary Topic
Hemoglobin structure and function
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Is Pulse Co-Oximetry a Reliable Alternative to Invasive Hemoglobin Measurement in Pediatric Neurosurgical Procedures?

Funda Arun, Oguzhan Arun
Children
Hemoglobin structure and function
article

Is Pulse Co-Oximetry a Reliable Alternative to Invasive Hemoglobin Measurement in Pediatric Neurosurgical Procedures?

Funda Arun, Oguzhan Arun
article en

Abstract

<b>Background/Objectives:</b> Pediatric neurosurgical procedures often involve significant blood loss and rapid hemodynamic shifts, necessitating accurate hemoglobin (Hb) monitoring. While continuous non-invasive Hb (SpHb) monitoring offers real-time trending, its accuracy in high-risk pediatric populations remains debated. We aimed to evaluate the diagnostic accuracy and clinical utility of SpHb compared to invasive arterial blood gas (ABG) analysis in pediatric patients undergoing cranial and spinal surgeries. <b>Methods:</b> This prospective, observational study enrolled 60 pediatric patients (aged 0-16 years) scheduled for high-risk neurosurgery. SpHb was measured continuously and compared with intermittent ABG-Hb values. Statistical analysis included Bland-Altman agreement, Pearson's correlation, and Error Grid Analysis. Subgroup analyses assessed the impact of the Perfusion Index (PI), hypotension, and metabolic acidosis on device performance. <b>Results:</b> Data from 57 patients (median age: 12 months, interquartile range: 6-42 months; 70.2% aged <24 months) were analyzed. SpHb demonstrated a moderate correlation with ABG-Hb (r = 0.567, <i>p</i> < 0.001) but exhibited systematic overestimation with a mean bias of +1.60 ± 1.54 g/dL. Crucially, SpHb showed 0% sensitivity for detecting critical anemia (Hb < 8.0 g/dL). Device performance was significantly compromised by physiological extremes: severe metabolic acidosis significantly increased bias to +2.27 g/dL (<i>p</i> = 0.038), and intraoperative hypotension significantly widened the limits of agreement (SD of bias: 1.79 g/dL vs. 1.45 g/dL in normotension). Furthermore, hemodynamic analysis revealed a loss of autoregulation during hypotension, where the pressure-perfusion coupling strengthened (r = 0.44) compared to the normotensive state (r = 0.15). <b>Conclusions:</b> SpHb monitoring provides fair Hb trending but is limited by systematic overestimation and poor sensitivity for critical anemia. Accuracy worsens during severe acidosis and hemodynamic instability. Therefore, SpHb should function as a complementary "early warning" trend monitor rather than a sole transfusion trigger, with invasive validation remaining essential for intraoperative decision-making.

ChildrenVol. 13(3)
Selçuk University (TR)
Good health and well-being
Openalex Percentile: Top 42%
Hemoglobin structure and function
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.