Variability in BIA-Derived Muscle Mass Estimates: Device Choice Impacts Diagnostic Classification

<i>Background/Objectives</i>: Although discrepancies between bioelectrical impedance analysis (BIA) devices are well documented, their clinical relevance in vulnerable populations remains unclear. This study aims to assess the impact of device choice on muscle mass classification criteria in patients with cancer or obesity and to identify modifiers of device variability. <i>Methods</i>: BIA data from 224 adults (85 with cancer, 139 with obesity) measured with two segmental multi-frequency devices (seca mBCA 515 and InBody 970) were analyzed. Device differences were assessed using the Wilcoxon signed-rank test and agreement analyses. Differences in classification of body composition cut-offs cited in the GLIM criteria for malnutrition and the ESPEN and EASO criteria for sarcopenic obesity were evaluated using McNemar's test. The impact of disease type, sex, and age on device differences was examined through multivariable models. <i>Results</i>: Significant device differences were found for all parameters (all <i>p</i> ≤ 0.005). Discrepancies were largest for skeletal muscle mass (kg and %), with effect sizes r > 0.8 and poor agreement (Lin's CCC < 0.90). A significant impact of device choice on muscle mass classification was observed for both cancer and obesity patients (<i>p</i> < 0.001), with seca classifying more patients as having low fat-free mass (50% vs. 20%) and as having a body composition consistent with sarcopenic obesity (90% vs. 50%) than InBody. Discrepancies were more pronounced in cancer patients and females. <i>Conclusions</i>: Muscle mass assessment by BIA is highly dependent on device choice, potentially leading to clinically relevant discrepancies in classification when rigid cut-offs are applied. An individualized interpretation of BIA data and further validation of prediction equations in disease-specific subpopulations is warranted.

Authors

Institutions

Publication Details

Journal
Nutrients
Published
2026-02-26
DOI
https://doi.org/10.3390/nu18050767
Citations
1
Primary Topic
Body Composition Measurement Techniques
Type
article
Field-Weighted Citation Impact
32.84

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Variability in BIA-Derived Muscle Mass Estimates: Device Choice Impacts Diagnostic Classification

Dejan Reljic, H. Herrmann, Verena Alexia Ohse, Yurdagül Herrmann Zopf et al.
1 citations
Nutrients
Body Composition Measurement Techniques
32.84
article

Variability in BIA-Derived Muscle Mass Estimates: Device Choice Impacts Diagnostic Classification

Dejan Reljic, H. Herrmann, Verena Alexia Ohse, Yurdagül Herrmann Zopf, Leonie Cordelia Burgard, Markus Friedrich Neurath, Siri Goldschmidt
article en
1 citations

Abstract

<i>Background/Objectives</i>: Although discrepancies between bioelectrical impedance analysis (BIA) devices are well documented, their clinical relevance in vulnerable populations remains unclear. This study aims to assess the impact of device choice on muscle mass classification criteria in patients with cancer or obesity and to identify modifiers of device variability. <i>Methods</i>: BIA data from 224 adults (85 with cancer, 139 with obesity) measured with two segmental multi-frequency devices (seca mBCA 515 and InBody 970) were analyzed. Device differences were assessed using the Wilcoxon signed-rank test and agreement analyses. Differences in classification of body composition cut-offs cited in the GLIM criteria for malnutrition and the ESPEN and EASO criteria for sarcopenic obesity were evaluated using McNemar's test. The impact of disease type, sex, and age on device differences was examined through multivariable models. <i>Results</i>: Significant device differences were found for all parameters (all <i>p</i> ≤ 0.005). Discrepancies were largest for skeletal muscle mass (kg and %), with effect sizes r > 0.8 and poor agreement (Lin's CCC < 0.90). A significant impact of device choice on muscle mass classification was observed for both cancer and obesity patients (<i>p</i> < 0.001), with seca classifying more patients as having low fat-free mass (50% vs. 20%) and as having a body composition consistent with sarcopenic obesity (90% vs. 50%) than InBody. Discrepancies were more pronounced in cancer patients and females. <i>Conclusions</i>: Muscle mass assessment by BIA is highly dependent on device choice, potentially leading to clinically relevant discrepancies in classification when rigid cut-offs are applied. An individualized interpretation of BIA data and further validation of prediction equations in disease-specific subpopulations is warranted.

NutrientsVol. 18(5)
Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Universitätsklinikum Erlangen (DE)
Friedrich-Alexander-Universität Erlangen-Nürnberg
Zero hunger
Openalex Percentile: Top 1%
Body Composition Measurement Techniques
32.84
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.