An Affordable and Simple Novel Approach to Obtain the Heat Transfer Coefficient (Kv) for Primary Drying Models Without Product Temperature Sensors

In this paper a novel method to obtain the heat transfer coefficient Kv value by the gravimetric method without using product temperature sensors is described. The sublimation front temperature is measured manometrically (indirectly by measuring the pressure). The key finding is a method to obtain the Kv value by using a primary drying model with a tailored algorithm. The authors have named it the gravi-manometric method. Kv values at different pressures are presented for two different containers: a freeze-drying stainless-steel cup, and for nested vials. The traditional gravimetric method has been compared against the new gravi-manometric method. Product temperature measurement can be challenging in applications with complex accessibility, as in large batch freeze dryers or when using protective membranes which enclose the product. This novel method is presented as an affordable and simple procedure by which the Kv value is obtained. The gravi-manometric method has the advantage that all the Kv values for a whole batch of vials can be obtained, as it is not limited to a discrete number of product temperature sensors.

Authors

Institutions

Publication Details

Journal
AAPS PharmSciTech
Published
2026-02-25
DOI
https://doi.org/10.1208/s12249-026-03328-3
Primary Topic
Food Drying and Modeling
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An Affordable and Simple Novel Approach to Obtain the Heat Transfer Coefficient (Kv) for Primary Drying Models Without Product Temperature Sensors

D. Samarkanova, M. Miñarro Carmona, J. R. Ticó, A. Nardi-Ricart et al.
AAPS PharmSciTech
Food Drying and Modeling
article

An Affordable and Simple Novel Approach to Obtain the Heat Transfer Coefficient (Kv) for Primary Drying Models Without Product Temperature Sensors

D. Samarkanova, M. Miñarro Carmona, J. R. Ticó, A. Nardi-Ricart, C. A. Amor
article en

Abstract

In this paper a novel method to obtain the heat transfer coefficient Kv value by the gravimetric method without using product temperature sensors is described. The sublimation front temperature is measured manometrically (indirectly by measuring the pressure). The key finding is a method to obtain the Kv value by using a primary drying model with a tailored algorithm. The authors have named it the gravi-manometric method. Kv values at different pressures are presented for two different containers: a freeze-drying stainless-steel cup, and for nested vials. The traditional gravimetric method has been compared against the new gravi-manometric method. Product temperature measurement can be challenging in applications with complex accessibility, as in large batch freeze dryers or when using protective membranes which enclose the product. This novel method is presented as an affordable and simple procedure by which the Kv value is obtained. The gravi-manometric method has the advantage that all the Kv values for a whole batch of vials can be obtained, as it is not limited to a discrete number of product temperature sensors.

AAPS PharmSciTechVol. 27(3)
Vall d'Hebron Institut de Recerca (ES), Ajuntament de L’Hospitalet (ES), Banc de Sang i Teixits (ES), Universitat de Barcelona (ES)
Openalex Percentile: Top 43%
Food Drying and Modeling
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.