Validation and performance assessment of a WD-XRF fused bead method for accurate quantification of scandium in nickel laterite deposits

This study focuses on the validation and performance assessment of the Wavelength Dispersive X-ray Fluorescence (WD-XRF) fused bead method for scandium quantification in nickel laterite ores. The calibration of the instrument showed a high correlation (R2 = 0.9962), with a Limit of Detection (LoD) of 20 ppm and a precision of 1.28% RSD. Accuracy was confirmed with good recoveries ranging from 98.05% to 101.46%. The relative expanded uncertainty was calculated at 5.89%. Method and instrumental reproducibility were confirmed by quality control (QC) Chart monitoring using certified reference materials (CRMs) and by relative error (RE) values below 10% for duplicate pulverized samples. The mode of scandium concentration in the samples was found to be 140 ppm. Overall, these results demonstrate that the WD-XRF fused bead method shows good validation performance, with excellent linearity, precision, accuracy, reproducibility, low expanded uncertainty and a low LoD for scandium quantification in complex matrices such as nickel laterite ores.

Authors

Publication Details

Journal
Springer Link (Chiba Institute of Technology)
Published
2026-02-24
DOI
https://doi.org/10.1051/e3sconf/202669502009/pdf
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation and performance assessment of a WD-XRF fused bead method for accurate quantification of scandium in nickel laterite deposits

Junjunan Muhammad Syukur
Springer Link (Chiba Institute of Technology)
X-ray Spectroscopy and Fluorescence Analysis
article

Validation and performance assessment of a WD-XRF fused bead method for accurate quantification of scandium in nickel laterite deposits

Junjunan Muhammad Syukur
article en

Abstract

This study focuses on the validation and performance assessment of the Wavelength Dispersive X-ray Fluorescence (WD-XRF) fused bead method for scandium quantification in nickel laterite ores. The calibration of the instrument showed a high correlation (R2 = 0.9962), with a Limit of Detection (LoD) of 20 ppm and a precision of 1.28% RSD. Accuracy was confirmed with good recoveries ranging from 98.05% to 101.46%. The relative expanded uncertainty was calculated at 5.89%. Method and instrumental reproducibility were confirmed by quality control (QC) Chart monitoring using certified reference materials (CRMs) and by relative error (RE) values below 10% for duplicate pulverized samples. The mode of scandium concentration in the samples was found to be 140 ppm. Overall, these results demonstrate that the WD-XRF fused bead method shows good validation performance, with excellent linearity, precision, accuracy, reproducibility, low expanded uncertainty and a low LoD for scandium quantification in complex matrices such as nickel laterite ores.

Springer Link (Chiba Institute of Technology)
Openalex Percentile: Top 38%
X-ray Spectroscopy and Fluorescence Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.