Exploring the Causes of Suture Needle Loss Through Simulation and Finite Element Analysis

The intraoperative loss of suture needles can prolong procedures and compromise safety, often occurring when a needle holder is unintentionally dropped and opens. This study explored the mechanism of needle loss through a nationwide survey, controlled drop simulations, and finite element analysis. Among 3649 OR nurses surveyed, 75.4% identified instrument transfer-specifically dropping the needle holder onto the instrument table-as the most common cause of loss. In 1500 simulated drops, unexpected opening occurred in 30.6% of cases, primarily when the holder landed flat or on its ring. Finite element analysis revealed that impact-induced clamp ring deformation displaces locking teeth beyond their biting depth, triggering release. Larger, heavier holders increased both opening risk and needle flight velocity. Design modifications, such as reinforcing the clamp ring and deepening locking teeth, along with cushioned surfaces, may reduce loss events and improve perioperative safety.

Authors

Publication Details

Journal
AORN Journal
Published
2026-02-25
DOI
https://doi.org/10.1002/aorn.70039
Primary Topic
Surgical Sutures and Adhesives
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Exploring the Causes of Suture Needle Loss Through Simulation and Finite Element Analysis

Lei Xiao Shuang, Yuhong Sun, Jing Wang, Ying Zhang et al.
AORN Journal
Surgical Sutures and Adhesives
article

Exploring the Causes of Suture Needle Loss Through Simulation and Finite Element Analysis

Lei Xiao Shuang, Yuhong Sun, Jing Wang, Ying Zhang, Wenxin Zhang
article en

Abstract

The intraoperative loss of suture needles can prolong procedures and compromise safety, often occurring when a needle holder is unintentionally dropped and opens. This study explored the mechanism of needle loss through a nationwide survey, controlled drop simulations, and finite element analysis. Among 3649 OR nurses surveyed, 75.4% identified instrument transfer-specifically dropping the needle holder onto the instrument table-as the most common cause of loss. In 1500 simulated drops, unexpected opening occurred in 30.6% of cases, primarily when the holder landed flat or on its ring. Finite element analysis revealed that impact-induced clamp ring deformation displaces locking teeth beyond their biting depth, triggering release. Larger, heavier holders increased both opening risk and needle flight velocity. Design modifications, such as reinforcing the clamp ring and deepening locking teeth, along with cushioned surfaces, may reduce loss events and improve perioperative safety.

AORN JournalVol. 123(3)
Openalex Percentile: Top 44%
Surgical Sutures and Adhesives
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.