Oxidised zirconium heads show a lower linear wear rate than 28-mm metal heads against XLPE: a retrospective analysis

There were no significant differences for any of the patient demographics between the 2 groups, with >60% women in both groups and normally distributed age. Cup inclination angle was also comparable between the 2 groups.While there was no difference in volumetric wear rates (<i>p</i> = 0.7377), the annual linear wear rate for OxZr was 0.030 mm (95% CI, 0.024-0.036) compared with 0.046 mm (95% CI, 0.038-0.054) for CoCr, a statistically significant difference (<i>p</i> = 0.0002)Conclusions:The widespread use of XLPE has revolutionised THA. This study is the first to show a statistically significantly lower wear rate with the use of OxZr. Whether this is enough to justify the cost differential and whether it translates to better implant survivorship is still to be determined. But if wear limiting wear rates is associated with decreasing revision rates, the use of ceramicised metal may be of clinical benefit.

Authors

Institutions

Publication Details

Journal
Hip International
Published
2026-02-25
DOI
https://doi.org/10.1177/11207000251412753
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Oxidised zirconium heads show a lower linear wear rate than 28-mm metal heads against XLPE: a retrospective analysis

Amit Atrey, Veronica Pentland, Richard W. McCalden, James P. Waddell et al.
Hip International
Orthopaedic implants and arthroplasty
article

Oxidised zirconium heads show a lower linear wear rate than 28-mm metal heads against XLPE: a retrospective analysis

Amit Atrey, Veronica Pentland, Richard W. McCalden, James P. Waddell, Amir Khoshbin, Zoe Thompson, Graeme Hoite, Douglas Naudie, Kevin Gargan
article en

Abstract

There were no significant differences for any of the patient demographics between the 2 groups, with >60% women in both groups and normally distributed age. Cup inclination angle was also comparable between the 2 groups.While there was no difference in volumetric wear rates (<i>p</i> = 0.7377), the annual linear wear rate for OxZr was 0.030 mm (95% CI, 0.024-0.036) compared with 0.046 mm (95% CI, 0.038-0.054) for CoCr, a statistically significant difference (<i>p</i> = 0.0002)Conclusions:The widespread use of XLPE has revolutionised THA. This study is the first to show a statistically significantly lower wear rate with the use of OxZr. Whether this is enough to justify the cost differential and whether it translates to better implant survivorship is still to be determined. But if wear limiting wear rates is associated with decreasing revision rates, the use of ceramicised metal may be of clinical benefit.

Hip International
Western University (CA), St. Michael's Hospital (CA), Nebraska Medical Center (US), University of Toronto (CA), London Health Sciences Centre (CA), University of Nebraska Medical Center (US)
Openalex Percentile: Top 44%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.