Performance Evaluation of AlphaSensor Radon Modules Under Real-World Conditions
This study compares a set of 43 AlphaSensor units produced by RadonTec GmbH, Wittislingen, Germany against the AlphaGUARD 1000PF (Bertin Technologies, Montigny-le-Bretonneux, France), which is used as a reference monitor. During this study, around 16 k integrated measurements were conducted. The concentration range varied between 10 and 20 k Bq/m<sup>3</sup>. Multiple key performance indicators, such as sensitivity, uncertainty, background, linearity, and temporal response, were evaluated using a variety of statistical approaches. The results confirm the manufacturer's claim of 10% or lower uncertainty in comparison with AlphaGUARD. We tentatively suggest individual calibration factors and methodologies. Our conclusion is that the AlphaSensor and commercial devices based on it, such as AlphaTracer, are affordable and applicable for home use. With modest additional hardware, AlphaSensors are also a good option for scientific studies involving the deployment of large monitoring networks.
Authors
- Stoyan Tenev
- L. Tsankov
- Atanas Terziyski (ORCID: https://orcid.org/0000-0001-7120-392X)
Institutions
- Plovdiv University (BG)
- Sofia University "St. Kliment Ohridski" (BG)
Publication Details
- Journal
- Sensors
- Published
- 2026-02-25
- DOI
- https://doi.org/10.3390/s26051432
- Primary Topic
- Radioactivity and Radon Measurements
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- European Commission