Performance evaluation of a flexible GAGG sheet for surface imaging of x-rays from computed tomography (CT) and linear accelerator (LINAC)

A flexible scintillator sheet made of Ce-doped Gd<sub>3</sub>Al<sub>2</sub>Ga<sub>3</sub>O<sub>12</sub>(GAGG) powder mixed with a two-part silicone adhesive enabled high-intensity imaging of proton beams on complex surfaces; however, its effectiveness for x-ray imaging remained unclear. To investigate this, we tested surface imaging with the GAGG sheet for both x-rays from computed tomography (CT) system and high-energy x-rays from linear accelerator (LINAC). For imaging of x-ray beam from CT system, we irradiated a flexible GAGG sheet with 140 kV x-rays, positioning it on a patient bed or curved surface of a cylindrical phantom. The resulting light emission on the GAGG sheet was captured by a CMOS camera. For LINAC x-ray imaging, we irradiated the GAGG sheet with 6 MV x-rays, setting it on a flat phantom or a human head phantom, and again captured the emitted light using a CMOS camera. In both CT and LINAC x-ray imaging, the light produced on the GAGG sheet was successfully captured at intervals as short as 100 ms, enabling real-time tracking of beams. The light output from CT x-rays was more than 50 times higher than that of plastic scintillator, while the light output from LINAC x-rays was 1.2 times higher. Given its adaptability to complex surfaces and high light emission for x-rays as well as real-time imaging capability, the flexible GAGG sheet shows potential for efficient surface beam imaging in both CT and LINAC x-ray applications.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2026-02-25
DOI
https://doi.org/10.1088/2057-1976/ae451e
Primary Topic
Digital Radiography and Breast Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Performance evaluation of a flexible GAGG sheet for surface imaging of x-rays from computed tomography (CT) and linear accelerator (LINAC)

Katsunori Yogo, Kohei Nakanishi, Kuniyasu Okudaira, Takayoshi Nakaya et al.
Biomedical Physics & Engineering Express
Digital Radiography and Breast Imaging
article

Performance evaluation of a flexible GAGG sheet for surface imaging of x-rays from computed tomography (CT) and linear accelerator (LINAC)

Katsunori Yogo, Kohei Nakanishi, Kuniyasu Okudaira, Takayoshi Nakaya, Masao Yoshino, Yumiko Noguchi, Seiichi Yamamoto, Jun Kataoka, Teiji Nishio, Kei Kamada, Akira Yoshikawa
article en

Abstract

A flexible scintillator sheet made of Ce-doped Gd<sub>3</sub>Al<sub>2</sub>Ga<sub>3</sub>O<sub>12</sub>(GAGG) powder mixed with a two-part silicone adhesive enabled high-intensity imaging of proton beams on complex surfaces; however, its effectiveness for x-ray imaging remained unclear. To investigate this, we tested surface imaging with the GAGG sheet for both x-rays from computed tomography (CT) system and high-energy x-rays from linear accelerator (LINAC). For imaging of x-ray beam from CT system, we irradiated a flexible GAGG sheet with 140 kV x-rays, positioning it on a patient bed or curved surface of a cylindrical phantom. The resulting light emission on the GAGG sheet was captured by a CMOS camera. For LINAC x-ray imaging, we irradiated the GAGG sheet with 6 MV x-rays, setting it on a flat phantom or a human head phantom, and again captured the emitted light using a CMOS camera. In both CT and LINAC x-ray imaging, the light produced on the GAGG sheet was successfully captured at intervals as short as 100 ms, enabling real-time tracking of beams. The light output from CT x-rays was more than 50 times higher than that of plastic scintillator, while the light output from LINAC x-rays was 1.2 times higher. Given its adaptability to complex surfaces and high light emission for x-rays as well as real-time imaging capability, the flexible GAGG sheet shows potential for efficient surface beam imaging in both CT and LINAC x-ray applications.

Biomedical Physics & Engineering ExpressVol. 12(2)
Waseda University (JP), Tohoku University (JP), Nagoya University Hospital (JP), Osaka Health Science University (JP), Nagoya University (JP)
Openalex Percentile: Top 44%
Digital Radiography and Breast Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.