A comparative study of serum electrolytes measured using flame photometry, direct ion-selective electrode method and indirect ion-selective electrode method

Abstract Background: Earlier studies showed differences in the values of serum electrolytes obtained by using different methods. In the present study three methods used for the measurement of serum electrolytes namely flame photometry (FP), direct ion-selective electrode method (ISE), and indirect ISE were compared to evaluate whether the values obtained by the methods are similar and interchangeable. Methods: Serum sodium and potassium levels were analyzed in 40 samples using the three methods. The United States Clinical Laboratory Improvement Amendments (US CLIA) 1988 guideline of acceptability of a difference of 0.5 mmol/L in potassium level and 4.0 mmol/L in sodium level compared to target values was considered as quality goal. Results: Significant difference was observed between paired values of all three methods of both electrolytes. The difference in sodium levels between FP and both ISE methods exceeded acceptable level while the difference in sodium levels between direct and indirect ISE methods is within the range given by US CLIA 88 guidelines. Serum potassium levels obtained from paired comparisons of the three methods was <0.5 mmol/L and hence was acceptable as per US CLIA 1988 guidelines. Significant positive correlation was found between electrolyte levels obtained by the three methods. There was a good agreement between direct ISE and indirect ISE methods for measurement of sodium. For potassium, there was a good agreement between paired values of all the three methods. Conclusions: Findings of present study indicate the importance of determining concordance between electrolyte values obtained using different methods. The differences that exist should be considered while interpreting the electrolyte levels.

Authors

Institutions

Publication Details

Journal
The Journal of Clinical and Scientific Research
Published
2025-10-01
DOI
https://doi.org/10.4103/jcsr.jcsr_82_24
Primary Topic
Renal function and acid-base balance
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparative study of serum electrolytes measured using flame photometry, direct ion-selective electrode method and indirect ion-selective electrode method

Sreekanth Pokala, V. S. Kiranmayi
The Journal of Clinical and Scientific Research
Renal function and acid-base balance
article

A comparative study of serum electrolytes measured using flame photometry, direct ion-selective electrode method and indirect ion-selective electrode method

Sreekanth Pokala, V. S. Kiranmayi
article en

Abstract

Abstract Background: Earlier studies showed differences in the values of serum electrolytes obtained by using different methods. In the present study three methods used for the measurement of serum electrolytes namely flame photometry (FP), direct ion-selective electrode method (ISE), and indirect ISE were compared to evaluate whether the values obtained by the methods are similar and interchangeable. Methods: Serum sodium and potassium levels were analyzed in 40 samples using the three methods. The United States Clinical Laboratory Improvement Amendments (US CLIA) 1988 guideline of acceptability of a difference of 0.5 mmol/L in potassium level and 4.0 mmol/L in sodium level compared to target values was considered as quality goal. Results: Significant difference was observed between paired values of all three methods of both electrolytes. The difference in sodium levels between FP and both ISE methods exceeded acceptable level while the difference in sodium levels between direct and indirect ISE methods is within the range given by US CLIA 88 guidelines. Serum potassium levels obtained from paired comparisons of the three methods was <0.5 mmol/L and hence was acceptable as per US CLIA 1988 guidelines. Significant positive correlation was found between electrolyte levels obtained by the three methods. There was a good agreement between direct ISE and indirect ISE methods for measurement of sodium. For potassium, there was a good agreement between paired values of all the three methods. Conclusions: Findings of present study indicate the importance of determining concordance between electrolyte values obtained using different methods. The differences that exist should be considered while interpreting the electrolyte levels.

The Journal of Clinical and Scientific ResearchVol. 14(4)
Sri Venkateswara Institute of Medical Sciences (IN)
Openalex Percentile: Top 25%
Renal function and acid-base balance
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.