Effective glycopeptide identification with tandem mass spectrometry (MS/MS) often relies on both low mass-to-charge (<i>m</i>/<i>z</i>) ions derived from glycan-specific oxonium ions and higher <i>m</i>/<i>z</i> peptide fragment ions that retain glycan modifications. Thus, glycoproteomic experiments benefit from a wider MS/MS scan range, i.e., the breadth of <i>m</i>/<i>z</i> values measured in fragmentation spectra, than those typically used in nonmodified peptide analyses. Here, we explore the implications of breaking a common axiom for scan range settings called the "5-10-15 rule" used to maximize transmission of ion populations of interest. This 5-10-15 rule, which defines the upper <i>m</i>/<i>z</i> value for a scan as a multiple of the first <i>m</i>/<i>z</i> value, comes from fundamental requirements for stable ion trajectories, where voltage settings must balance retention of low <i>m</i>/<i>z</i> ions while also generating effective pseudopotential wells to trap high <i>m</i>/<i>z</i> ions. Adhering to this calculation for MS/MS scan range settings can reduce glycopeptide ion coverage by excluding the analysis of either low <i>m</i>/<i>z</i> oxonium ions or high <i>m</i>/<i>z</i> fragment ions. We use a quadrupole-Orbitrap-linear ion trap Tribrid MS system (Orbitrap Ascend) to investigate the implications of following or breaking the 5-10-15 rule in MS/MS scans for glycopeptide characterization with higher-energy collisional dissociation (HCD), electron-transfer dissociation (ETD), and electron-transfer/higher-energy collision dissociation (EThcD). For scans with a first <i>m</i>/<i>z</i> value around <i>m</i>/<i>z</i> 120 (i.e., capturing most common glycan-specific oxonium ions), we show that breaking the 5-10-15 rule does not lead to a significant loss of fragment ion transmission at either extreme of the <i>m</i>/<i>z</i> range. We use this case study to discuss the concepts important to using the 5-10-15 rule wisely and when it can be practically ignored, such as using large scan ranges to improve glycopeptide characterization.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2026-02-23
DOI
https://doi.org/10.1021/jasms.5c00352
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Understanding m / z Range Settings for MS/MS Scans: A Case Study with Intact Glycopeptides

Nicholas M. Riley, Tim S. Veth, Kathryn Kothlow
Journal of the American Society for Mass Spectrometry
Mass Spectrometry Techniques and Applications
article

Understanding m / z Range Settings for MS/MS Scans: A Case Study with Intact Glycopeptides

Nicholas M. Riley, Tim S. Veth, Kathryn Kothlow
article en

Abstract

Effective glycopeptide identification with tandem mass spectrometry (MS/MS) often relies on both low mass-to-charge (<i>m</i>/<i>z</i>) ions derived from glycan-specific oxonium ions and higher <i>m</i>/<i>z</i> peptide fragment ions that retain glycan modifications. Thus, glycoproteomic experiments benefit from a wider MS/MS scan range, i.e., the breadth of <i>m</i>/<i>z</i> values measured in fragmentation spectra, than those typically used in nonmodified peptide analyses. Here, we explore the implications of breaking a common axiom for scan range settings called the "5-10-15 rule" used to maximize transmission of ion populations of interest. This 5-10-15 rule, which defines the upper <i>m</i>/<i>z</i> value for a scan as a multiple of the first <i>m</i>/<i>z</i> value, comes from fundamental requirements for stable ion trajectories, where voltage settings must balance retention of low <i>m</i>/<i>z</i> ions while also generating effective pseudopotential wells to trap high <i>m</i>/<i>z</i> ions. Adhering to this calculation for MS/MS scan range settings can reduce glycopeptide ion coverage by excluding the analysis of either low <i>m</i>/<i>z</i> oxonium ions or high <i>m</i>/<i>z</i> fragment ions. We use a quadrupole-Orbitrap-linear ion trap Tribrid MS system (Orbitrap Ascend) to investigate the implications of following or breaking the 5-10-15 rule in MS/MS scans for glycopeptide characterization with higher-energy collisional dissociation (HCD), electron-transfer dissociation (ETD), and electron-transfer/higher-energy collision dissociation (EThcD). For scans with a first <i>m</i>/<i>z</i> value around <i>m</i>/<i>z</i> 120 (i.e., capturing most common glycan-specific oxonium ions), we show that breaking the 5-10-15 rule does not lead to a significant loss of fragment ion transmission at either extreme of the <i>m</i>/<i>z</i> range. We use this case study to discuss the concepts important to using the 5-10-15 rule wisely and when it can be practically ignored, such as using large scan ranges to improve glycopeptide characterization.

Journal of the American Society for Mass Spectrometry
University of Washington (US)
Affordable and clean energy
Openalex Percentile: Top 46%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.