Comparative Dosimetric Analysis of System Displayed versus Calculated Entrance Skin Dose in Routine Radiographic Imaging

Abstract X-ray examinations are integral to medical diagnostics but pose risks from ionizing radiation. Adhering to the ALARA (as low as reasonably achievable) principle in diagnostic radiography is essential. This study evaluates the accuracy and reliability of entrance skin dose (ESD) values displayed on digital radiographic X-ray machine by comparing them with estimated ESD values during common radiographic examinations. ESD was estimated for 250 radiographs of 150 patients (90 females, 60 males) undergoing five common radiographic views: lumbosacral (LS) spine anterior–posterior (AP)/lateral (LAT), pelvis AP, and abdomen supine/erect. Using the formula, estimated ESD values were calculated and compared with displayed ESD values from the console of the digital X-ray machine. The mean estimated and displayed ESD values were: LS spine AP (1.31 mGy, 0.92 mGy), LS spine LAT (2.12 mGy, 1.63 mGy), pelvis AP (0.69 mGy, 0.62 mGy), abdomen supine (0.67 mGy, 0.65 mGy), and abdomen erect (1.44 mGy, 1.06 mGy). Statistical analysis revealed a significant correlation between estimated and displayed ESD values across all views (p <0.001). The intraclass correlation coefficient ranged from 62.60 to 84.30%, indicating moderate to strong agreement. These findings were compared with studies from Canada, India, and the UK. Displayed ESD values showed significant deviation from the independently estimated ESD values, indicating that console-generated estimates provide only an approximate indication of patient exposure. Therefore, they should not be considered a precise substitute for direct dosimetric measurements. Reliable ESD values displayed on radiographic equipment facilitate effective radiation protection measures. Incorporating regular ESD monitoring into clinical practice can enhance patient safety and adherence to radiation safety protocols in diagnostic radiology.

Authors

Institutions

Publication Details

Journal
Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging
Published
2026-02-24
DOI
https://doi.org/10.1055/s-0046-1817157
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Dosimetric Analysis of System Displayed versus Calculated Entrance Skin Dose in Routine Radiographic Imaging

AK Srivastava, Atul Mishra, Abhishek Kumar, Shuchi Bhatt
Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging
Radiation Dose and Imaging
article

Comparative Dosimetric Analysis of System Displayed versus Calculated Entrance Skin Dose in Routine Radiographic Imaging

AK Srivastava, Atul Mishra, Abhishek Kumar, Shuchi Bhatt
article en

Abstract

Abstract X-ray examinations are integral to medical diagnostics but pose risks from ionizing radiation. Adhering to the ALARA (as low as reasonably achievable) principle in diagnostic radiography is essential. This study evaluates the accuracy and reliability of entrance skin dose (ESD) values displayed on digital radiographic X-ray machine by comparing them with estimated ESD values during common radiographic examinations. ESD was estimated for 250 radiographs of 150 patients (90 females, 60 males) undergoing five common radiographic views: lumbosacral (LS) spine anterior–posterior (AP)/lateral (LAT), pelvis AP, and abdomen supine/erect. Using the formula, estimated ESD values were calculated and compared with displayed ESD values from the console of the digital X-ray machine. The mean estimated and displayed ESD values were: LS spine AP (1.31 mGy, 0.92 mGy), LS spine LAT (2.12 mGy, 1.63 mGy), pelvis AP (0.69 mGy, 0.62 mGy), abdomen supine (0.67 mGy, 0.65 mGy), and abdomen erect (1.44 mGy, 1.06 mGy). Statistical analysis revealed a significant correlation between estimated and displayed ESD values across all views (p <0.001). The intraclass correlation coefficient ranged from 62.60 to 84.30%, indicating moderate to strong agreement. These findings were compared with studies from Canada, India, and the UK. Displayed ESD values showed significant deviation from the independently estimated ESD values, indicating that console-generated estimates provide only an approximate indication of patient exposure. Therefore, they should not be considered a precise substitute for direct dosimetric measurements. Reliable ESD values displayed on radiographic equipment facilitate effective radiation protection measures. Incorporating regular ESD monitoring into clinical practice can enhance patient safety and adherence to radiation safety protocols in diagnostic radiology.

Indian journal of radiology and imaging - new series/Indian journal of radiology and imaging/Indian Journal of Radiology & Imaging
University College of Medical Sciences (IN), Sarojini Naidu Medical College (IN), Adesh University (IN), All India Institute of Medical Sciences (IN)
Good health and well-being
Openalex Percentile: Top 45%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.