Organ Dose Assessment in Radiological Central Venous Port Procedures: An Experimental Phantom-Based Study

This study aimed to quantitatively evaluate radiation doses delivered to radiosensitive thoracic and cervical organs during fluoroscopy-guided central venous port (CVP) catheter placement using a standardized experimental setup to identify the most exposed anatomical regions. An Alderson Rando anthropomorphic phantom equipped with calibrated MTS-100 thermoluminescent dosimeters (TLDs) was employed to simulate a typical CVP fluoroscopic procedure. The setup used a Siemens Artis Zee C-arm fluoroscopy system with a total active exposure time of 8 min, replicating clinical practice. Absorbed dose measurements were obtained from 14 organ sites, including lungs, thyroid, spinal cord, and upper airway structures. The highest radiation doses were observed in thoracic organs, with measured values of 7.84 ± 0.28 mGy for the right middle lung, 5.23 ± 0.40 mGy for the thoracic spinal cord, and 3.91 ± 1.47 mGy for the right lung apex. The thyroid received moderate exposure, recording 1.48 ± 0.35 mGy (right lobe) and 1.41 ± 0.68 mGy (left lobe). In contrast, distant organs such as the pituitary gland and ethmoid sinus exhibited negligible absorbed doses (< 0.03mGy). Fluoroscopic CVP placement results in substantial localized radiation exposure, particularly to the lungs and thyroid. Quantitative organ dose mapping using an anthropomorphic phantom provides a valuable reference for optimizing patient safety protocols and minimizing radiation exposure during interventional procedures.

Authors

Institutions

Publication Details

Journal
Health Physics
Published
2026-02-24
DOI
https://doi.org/10.1097/hp.0000000000002138
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Organ Dose Assessment in Radiological Central Venous Port Procedures: An Experimental Phantom-Based Study

HO Tekin, Songül Çavdar Karaçam, Mehmet Toptaş, Özlem Ateşal et al.
Health Physics
Radiation Dose and Imaging
article

Organ Dose Assessment in Radiological Central Venous Port Procedures: An Experimental Phantom-Based Study

HO Tekin, Songül Çavdar Karaçam, Mehmet Toptaş, Özlem Ateşal, Osman Günay, Abdurrahman Tünay, Mustafa Demir, DC Oksüz, D. Tunçman Kayaokay, Ghada Al Misned, Fahrettin Fatih Kesmezacar
article en

Abstract

This study aimed to quantitatively evaluate radiation doses delivered to radiosensitive thoracic and cervical organs during fluoroscopy-guided central venous port (CVP) catheter placement using a standardized experimental setup to identify the most exposed anatomical regions. An Alderson Rando anthropomorphic phantom equipped with calibrated MTS-100 thermoluminescent dosimeters (TLDs) was employed to simulate a typical CVP fluoroscopic procedure. The setup used a Siemens Artis Zee C-arm fluoroscopy system with a total active exposure time of 8 min, replicating clinical practice. Absorbed dose measurements were obtained from 14 organ sites, including lungs, thyroid, spinal cord, and upper airway structures. The highest radiation doses were observed in thoracic organs, with measured values of 7.84 ± 0.28 mGy for the right middle lung, 5.23 ± 0.40 mGy for the thoracic spinal cord, and 3.91 ± 1.47 mGy for the right lung apex. The thyroid received moderate exposure, recording 1.48 ± 0.35 mGy (right lobe) and 1.41 ± 0.68 mGy (left lobe). In contrast, distant organs such as the pituitary gland and ethmoid sinus exhibited negligible absorbed doses (< 0.03mGy). Fluoroscopic CVP placement results in substantial localized radiation exposure, particularly to the lungs and thyroid. Quantitative organ dose mapping using an anthropomorphic phantom provides a valuable reference for optimizing patient safety protocols and minimizing radiation exposure during interventional procedures.

Health Physics
Princess Nourah bint Abdulrahman University (SA), University of Sharjah (AE), Yıldız Technical University (TR), Istanbul University-Cerrahpaşa (TR), Istinye University (TR), Sağlık Bilimleri Üniversitesi (TR), Türk Anesteziyoloji ve Reanimasyon Derneği (TR), Istanbul University (TR), University of Health Sciences Antigua (AG)
Good health and well-being
Openalex Percentile: Top 46%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.