The MAIIA EPO Purification Gel Kit, 7D3—An Alternative Purification Kit for Screening and Confirmation Procedures. Validation for Serum and Plasma Matrices

In this study, the performance of the MAIIA EPO purification kit 7D3 was evaluated for human serum and plasma matrices. The kit was validated for selectivity in comparison to the MAIIA 3F6 kit, reliability at minimum required performance levels (MRPLs), limit of detection (LOD), carryover and recovery to meet compliance with the World Anti-Doping Agency criteria of the technical document 2024 (TD2024EPO). In addition, the influence of haemolysis on plasma samples was evaluated. In total, 30 serum samples and 30 plasma samples were tested for selectivity; 10 serum and 10 plasma samples, for reliability at MRPL and recovery; and six serum and six plasma samples, for LOD. The influence of haemolysis was evaluated by creating mixtures of 0%, 25%, 50%, 75% and 100% of a hemolytic sample and a blank sample, and the same sample spiked at the MRPL level. This study validates that the 7D3 kit meets all WADA TD2024EPO criteria, providing a reliable alternative for the purification of erythropoietin receptor agonists from human serum and plasma matrices in both screening and confirmation procedures in routine antidoping analysis.

Authors

Institutions

Publication Details

Journal
Drug Testing and Analysis
Published
2026-02-24
DOI
https://doi.org/10.1002/dta.70051
Primary Topic
Erythropoietin and Anemia Treatment
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The MAIIA EPO Purification Gel Kit, 7D3—An Alternative Purification Kit for Screening and Confirmation Procedures. Validation for Serum and Plasma Matrices

Jana Hempel, Dirk Schwenke, S Reimann, S C Voss
Drug Testing and Analysis
Erythropoietin and Anemia Treatment
article

The MAIIA EPO Purification Gel Kit, 7D3—An Alternative Purification Kit for Screening and Confirmation Procedures. Validation for Serum and Plasma Matrices

Jana Hempel, Dirk Schwenke, S Reimann, S C Voss
article en

Abstract

In this study, the performance of the MAIIA EPO purification kit 7D3 was evaluated for human serum and plasma matrices. The kit was validated for selectivity in comparison to the MAIIA 3F6 kit, reliability at minimum required performance levels (MRPLs), limit of detection (LOD), carryover and recovery to meet compliance with the World Anti-Doping Agency criteria of the technical document 2024 (TD2024EPO). In addition, the influence of haemolysis on plasma samples was evaluated. In total, 30 serum samples and 30 plasma samples were tested for selectivity; 10 serum and 10 plasma samples, for reliability at MRPL and recovery; and six serum and six plasma samples, for LOD. The influence of haemolysis was evaluated by creating mixtures of 0%, 25%, 50%, 75% and 100% of a hemolytic sample and a blank sample, and the same sample spiked at the MRPL level. This study validates that the 7D3 kit meets all WADA TD2024EPO criteria, providing a reliable alternative for the purification of erythropoietin receptor agonists from human serum and plasma matrices in both screening and confirmation procedures in routine antidoping analysis.

Drug Testing and Analysis
Institute for Sports Medicine (DE), Technische Universität Dresden (DE)
Good health and well-being
Openalex Percentile: Top 45%
Erythropoietin and Anemia Treatment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.