Installation, Commissioning, and Performance Evaluation of Bhabhatron‑II TAW – An Indian‑Made Telecobalt Unit

Aim: The aim of this study was to report commissioning and performance evaluation of BhabhatronII TAW (an Indian made machine). Materials and Methods: Commissioning comprises of following quality assurance (QA) (a) electrical QA, (b) mechanical QA, (c) radiation QA and (d) survey of installation. Electrical QA were done for various components of the unit for its functionality. Mechanical QA were done and the measured data were verified as per guidelines of Atomic Energy Regulatory Board (AERB), India and vendor’s specified values. Cylindrical ionization chamber, electrometer, ready pack films, water phantom, thermometer and barometer were used during radiation QA and survey of installation procedures. Moreover, the data so generated were fed to treatment planning system (TPS) for dose calculation and also kept in proper tabular manner for clinical use. Results: Couch and collimator parameters were found to comply the tolerance criteria. Source strength of 172.2 roentgen per minute at 1 meter (RMM) was measured with an absolute dose rate of 225.76 cGy/min at dmax for 10cm x 10cm field size at normal treatment distance (NTD). Shutter timer error of 0.004 min was measured and incorporated in the clinical data. A maximum of 3.66 % variation was noted for 300 (15W × 20) physical wedge transmission factor between measured and vendor specified value. Radiation leakage through beam limiting device (BLD) and outside BLD were found to be within the limits prescribed by AERB. Conclusion: Commissioning is a prerequisite process after installation of a radiation therapy machine for medical purpose. Bhabhatron-II machine has several advantages over other telecobalt machine. Features like high source storage capacity of machine head, one pair of asymmetric jaw, closure of jaws to 0 cm x 0 cm during radiation emergency and presence of one motorized wedge filter makes it unique and yet cost effective telecobalt machine..

Authors

Institutions

Publication Details

Published
2026-02-12
DOI
https://doi.org/10.82339/jrmt.vol1no1.64
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Installation, Commissioning, and Performance Evaluation of Bhabhatron‑II TAW – An Indian‑Made Telecobalt Unit

Bhaveshwar Yadav, Shachindra Goswami, Gautam Sarma, Mithu Barthakur et al.
Advanced Radiotherapy Techniques
article

Installation, Commissioning, and Performance Evaluation of Bhabhatron‑II TAW – An Indian‑Made Telecobalt Unit

Bhaveshwar Yadav, Shachindra Goswami, Gautam Sarma, Mithu Barthakur, Pranjal Goswami, Shashi Bhushan Sharma
article en

Abstract

Aim: The aim of this study was to report commissioning and performance evaluation of BhabhatronII TAW (an Indian made machine). Materials and Methods: Commissioning comprises of following quality assurance (QA) (a) electrical QA, (b) mechanical QA, (c) radiation QA and (d) survey of installation. Electrical QA were done for various components of the unit for its functionality. Mechanical QA were done and the measured data were verified as per guidelines of Atomic Energy Regulatory Board (AERB), India and vendor’s specified values. Cylindrical ionization chamber, electrometer, ready pack films, water phantom, thermometer and barometer were used during radiation QA and survey of installation procedures. Moreover, the data so generated were fed to treatment planning system (TPS) for dose calculation and also kept in proper tabular manner for clinical use. Results: Couch and collimator parameters were found to comply the tolerance criteria. Source strength of 172.2 roentgen per minute at 1 meter (RMM) was measured with an absolute dose rate of 225.76 cGy/min at dmax for 10cm x 10cm field size at normal treatment distance (NTD). Shutter timer error of 0.004 min was measured and incorporated in the clinical data. A maximum of 3.66 % variation was noted for 300 (15W × 20) physical wedge transmission factor between measured and vendor specified value. Radiation leakage through beam limiting device (BLD) and outside BLD were found to be within the limits prescribed by AERB. Conclusion: Commissioning is a prerequisite process after installation of a radiation therapy machine for medical purpose. Bhabhatron-II machine has several advantages over other telecobalt machine. Features like high source storage capacity of machine head, one pair of asymmetric jaw, closure of jaws to 0 cm x 0 cm during radiation emergency and presence of one motorized wedge filter makes it unique and yet cost effective telecobalt machine..

Bhubaneswar Borooah Cancer Institute (IN)
Affordable and clean energy
Openalex Percentile: Top 41%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.