Increased Skin-to-Muscle Distances Affects the Dose of Adrenaline Administered Intramuscularly From EpiPen With Certain Cold Weather Clothing Combinations

IntroductionIn extreme cold weather (ECW) environments it may be beneficial to activate an auto-injector through clothing rather than against skin to minimize casualty exposure. It has been shown that popular adrenaline auto-injectors may not deliver an adequate intramuscular dose to individuals with increased subcutaneous fat, even when activated against skin. This experiment sought to build on research into auto-injector use in the ECW environment by assessing drug delivery through winter clothing combinations with varying simulated skin-to-muscle distances (STMD).MethodologyInvestigators based at the British Antarctic Survey Medical Unit (BASMU) in Plymouth, UK, tested two brands of auto-injector, EpiPen and Emerade, through commonly worn clothing sourced from the British Antarctic Survey. The experiment involved activating the auto-injector through clothing layers and different depths of silicone representing subcutaneous tissue. The volume administered through these layers was collected and its mass was measured. This mass was taken as a representation of the drug volume delivered through the various clothing layers, subcutaneous tissue, and muscle.ResultsThe EpiPen administered adrenaline through all clothing combinations tested at STMD 5 mm; however, thicker clothing combinations at STMD 10 mm prevented effective drug delivery. The Emerade auto-injectors delivered adrenaline at an acceptable dose through the thicker combinations at STMD of 10 mm and 15 mm.ConclusionsThese results suggest that adrenaline auto-injectors can successfully activate through polar clothing but that the depth of injection may be reduced. This may be more consequential in individuals with a greater STMD and should be considered when treating patients in cold environments.

Authors

Institutions

Publication Details

Journal
Wilderness and Environmental Medicine
Published
2026-02-23
DOI
https://doi.org/10.1177/10806032251413551
Primary Topic
Infection Control and Ventilation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Increased Skin-to-Muscle Distances Affects the Dose of Adrenaline Administered Intramuscularly From EpiPen With Certain Cold Weather Clothing Combinations

Jonathon Lowe, Timothy Parham, Felix NR Wood, Tim D Moloney
Wilderness and Environmental Medicine
Infection Control and Ventilation
article

Increased Skin-to-Muscle Distances Affects the Dose of Adrenaline Administered Intramuscularly From EpiPen With Certain Cold Weather Clothing Combinations

Jonathon Lowe, Timothy Parham, Felix NR Wood, Tim D Moloney
article en

Abstract

IntroductionIn extreme cold weather (ECW) environments it may be beneficial to activate an auto-injector through clothing rather than against skin to minimize casualty exposure. It has been shown that popular adrenaline auto-injectors may not deliver an adequate intramuscular dose to individuals with increased subcutaneous fat, even when activated against skin. This experiment sought to build on research into auto-injector use in the ECW environment by assessing drug delivery through winter clothing combinations with varying simulated skin-to-muscle distances (STMD).MethodologyInvestigators based at the British Antarctic Survey Medical Unit (BASMU) in Plymouth, UK, tested two brands of auto-injector, EpiPen and Emerade, through commonly worn clothing sourced from the British Antarctic Survey. The experiment involved activating the auto-injector through clothing layers and different depths of silicone representing subcutaneous tissue. The volume administered through these layers was collected and its mass was measured. This mass was taken as a representation of the drug volume delivered through the various clothing layers, subcutaneous tissue, and muscle.ResultsThe EpiPen administered adrenaline through all clothing combinations tested at STMD 5 mm; however, thicker clothing combinations at STMD 10 mm prevented effective drug delivery. The Emerade auto-injectors delivered adrenaline at an acceptable dose through the thicker combinations at STMD of 10 mm and 15 mm.ConclusionsThese results suggest that adrenaline auto-injectors can successfully activate through polar clothing but that the depth of injection may be reduced. This may be more consequential in individuals with a greater STMD and should be considered when treating patients in cold environments.

Wilderness and Environmental Medicine
British Antarctic Survey (GB), Royal Air Force College Cranwell (GB), Royal Navy (GB)
Good health and well-being
Openalex Percentile: Top 46%
Infection Control and Ventilation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.