A comparative evaluation of primary stability of two dental implant systems in synthetic and natural bone

Given the increasing use and availability of dental implants, this study aimed to compare the primary stability of two commonly used implants with similar designs: 'type A (Cowellmedi), an imported design, and type B (DRI), a domestically manufactured implant,' in both synthetic (polyurethane foam) and natural bone (bovine rib). Both implants had a diameter of 4 mm and a length of 10 mm, with a buttress thread profile. Type B was single-threaded, while type A was double-threaded with a more conical design. In natural bone, type A demonstrated significantly higher maximum insertion torque (<i>p</i> = 0.033), bone-implant construct stiffness (<i>p</i> < 0.001), and maximum push-in force (<i>p</i> < 0.001) compared with type B. In contrast, in synthetic bone, type B showed significantly higher bone-implant construct stiffness (<i>p</i> = 0.02) and greater maximum push-in force (<i>p</i> < 0.001). Implant stability quotient (ISQ) values did not differ between the two designs in either bone type. The findings suggest that type A achieved greater primary stability due to its double-threaded configuration and longer conical region. Subtle design differences, such as single- versus double-threading and a longer conical region, can therefore have a considerable impact on primary stability and should not be overlooked when developing new implant designs. Finally, the discrepancies observed between stability outcomes in natural and synthetic bone are likely due to differences in their mechanical behavior. This highlights the need for caution when interpreting data derived from synthetic bone, particularly when attempting to extrapolate findings to natural bone-implant systems.

Authors

Institutions

Publication Details

Journal
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Published
2026-02-22
DOI
https://doi.org/10.1177/09544119251414193
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A comparative evaluation of primary stability of two dental implant systems in synthetic and natural bone

Tahereh Mirzaie, Amirreza Rokn, Mahdi Farrokhi Kashtiban, Armin Shanazari et al.
Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in Medicine
Dental Implant Techniques and Outcomes
article

A comparative evaluation of primary stability of two dental implant systems in synthetic and natural bone

Tahereh Mirzaie, Amirreza Rokn, Mahdi Farrokhi Kashtiban, Armin Shanazari, Gholamreza Rouhi
article en

Abstract

Given the increasing use and availability of dental implants, this study aimed to compare the primary stability of two commonly used implants with similar designs: 'type A (Cowellmedi), an imported design, and type B (DRI), a domestically manufactured implant,' in both synthetic (polyurethane foam) and natural bone (bovine rib). Both implants had a diameter of 4 mm and a length of 10 mm, with a buttress thread profile. Type B was single-threaded, while type A was double-threaded with a more conical design. In natural bone, type A demonstrated significantly higher maximum insertion torque (<i>p</i> = 0.033), bone-implant construct stiffness (<i>p</i> < 0.001), and maximum push-in force (<i>p</i> < 0.001) compared with type B. In contrast, in synthetic bone, type B showed significantly higher bone-implant construct stiffness (<i>p</i> = 0.02) and greater maximum push-in force (<i>p</i> < 0.001). Implant stability quotient (ISQ) values did not differ between the two designs in either bone type. The findings suggest that type A achieved greater primary stability due to its double-threaded configuration and longer conical region. Subtle design differences, such as single- versus double-threading and a longer conical region, can therefore have a considerable impact on primary stability and should not be overlooked when developing new implant designs. Finally, the discrepancies observed between stability outcomes in natural and synthetic bone are likely due to differences in their mechanical behavior. This highlights the need for caution when interpreting data derived from synthetic bone, particularly when attempting to extrapolate findings to natural bone-implant systems.

Proceedings of the Institution of Mechanical Engineers Part H Journal of Engineering in MedicineVol. 240(3)
Amirkabir University of Technology (IR), Iran University of Medical Sciences (IR), Tehran University of Medical Sciences (IR)
Openalex Percentile: Top 48%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.