When cold complicates: the impact of cryoglobulins and pre-analytical conditions on monoclonal protein analysis

Cryoprecipitation of monoclonal protein in two patients from separate hospitals led to falsely reduced concentrations. In one case, this was only later discovered because the hematology analyzer falsely flagged the cryoglobulins as abnormal platelets and lymphocytosis. This resulted in the generation of a peripheral blood smear in which the cryoprecipitation was microscopically identified. In both cases, initially unnoticed discrepancies existed between the total immunoglobulin and albumin concentrations derived from the general chemistry analyzer and those from serum protein electrophoresis analysis, which is what later alerted staff to the second case. These cases highlight the significant risk cryoglobulins pose in the diagnosis and accurate follow-up of plasma cell dyscrasias. We therefore propose several measures to improve laboratory detection and prevention of such discrepancies. These include; performing total immunoglobulin and albumin measurements alongside the protein electrophoresis analysis; implementation of a cross reference alarm in the laboratory information system for the discrepancies; consequent visual inspection of all samples prior to analysis; pre-warming of samples with a visible cryoprecipitate prior to serum protein electrophoresis; and the implementation of adequate operating procedures for handling samples from patients with known cryoglobulins to prevent pre-analytical loss caused by precipitation. Finally, we provide insight into the stability of immunoglobulins and monoclonal proteins at room temperature, which may be helpful to determine the optimal sample storage temperature in order to reduce the risk of unwanted cryoprecipitation.

Authors

Institutions

Publication Details

Journal
Clinical and Experimental Medicine
Published
2026-02-22
DOI
https://doi.org/10.1007/s10238-026-02088-5
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

When cold complicates: the impact of cryoglobulins and pre-analytical conditions on monoclonal protein analysis

André Vlot, Jolanda Scherrenburg, Joannes F.M. Jacobs, Monique Van Uum-Otters et al.
Clinical and Experimental Medicine
Clinical Laboratory Practices and Quality Control
article

When cold complicates: the impact of cryoglobulins and pre-analytical conditions on monoclonal protein analysis

André Vlot, Jolanda Scherrenburg, Joannes F.M. Jacobs, Monique Van Uum-Otters, Dorien M. Rotteveel, Marlies Oostendorp, Dennis van Aggelen, Rebecca Verheggen, Janneke Ruinemans-Koerts, Alan Cunningham
article en

Abstract

Cryoprecipitation of monoclonal protein in two patients from separate hospitals led to falsely reduced concentrations. In one case, this was only later discovered because the hematology analyzer falsely flagged the cryoglobulins as abnormal platelets and lymphocytosis. This resulted in the generation of a peripheral blood smear in which the cryoprecipitation was microscopically identified. In both cases, initially unnoticed discrepancies existed between the total immunoglobulin and albumin concentrations derived from the general chemistry analyzer and those from serum protein electrophoresis analysis, which is what later alerted staff to the second case. These cases highlight the significant risk cryoglobulins pose in the diagnosis and accurate follow-up of plasma cell dyscrasias. We therefore propose several measures to improve laboratory detection and prevention of such discrepancies. These include; performing total immunoglobulin and albumin measurements alongside the protein electrophoresis analysis; implementation of a cross reference alarm in the laboratory information system for the discrepancies; consequent visual inspection of all samples prior to analysis; pre-warming of samples with a visible cryoprecipitate prior to serum protein electrophoresis; and the implementation of adequate operating procedures for handling samples from patients with known cryoglobulins to prevent pre-analytical loss caused by precipitation. Finally, we provide insight into the stability of immunoglobulins and monoclonal proteins at room temperature, which may be helpful to determine the optimal sample storage temperature in order to reduce the risk of unwanted cryoprecipitation.

Clinical and Experimental MedicineVol. 26(1)
Radboud University Nijmegen (NL), Radboud University Medical Center (NL), Canisius-Wilhelmina Ziekenhuis (NL), Rijnstate Hospital (NL)
Openalex Percentile: Top 46%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.