Evaluation of a research use only (RUO) TIMP-1 assay for use with the automated Alinity i platform

Tissue inhibitor of metalloproteinase-1 (TIMP-1) binds matrix metalloproteinases to regulate extracellular matrix turnover; elevated TIMP-1 has been associated with liver fibrosis, tumor progression, and fibrotic heart disease. We previously developed a quantitative, research use only (RUO) automated TIMP-1 immunoassay for use with the Abbott ARCHITECT <i>i</i> System; here, we describe the modification of the assay for use with the Abbott Alinity <i>i</i> system. The observed limit of detection was 1.42 ng/mL, and the calculated limit of quantitation was 2.44 ng/mL. In 100 normal plasma samples, TIMP-1 levels ranged from 106.23 ng/mL to 329.68 ng/mL. Deviation from linearity was ≤10% and precision was ≤10%CV across the analytical measuring interval from 2.44 to 500 ng/mL. No significant interference was observed for both conjugated and unconjugated bilirubin and hemoglobin at industry recommended test concentrations. Slight interference for protein and triglyceride resolved with titration to 12.15 g/dL and 750 mg/dL, respectively. Assay performance was optimal for plasma specimens in LiHep or EDTA. TIMP-1 in plasma samples was stable after multiple freeze-thaw cycles, up to 7 days at 2‒8 °C, and for up to 3 h on-board the Alinity <i>i</i> instrument. In cardiac plasma specimens, the assay consistently detected higher TIMP-1 in specimens with elevated troponin-I. The TIMP-1 immunoassay, modified for the Alinity<i>i</i> platform, detected TIMP-1 at physiologically relevant ranges. Future clinical development of the assay will focus on its clinical utility in various cardiac cohorts as a potential biomarker of fibrotic heart disease.

Authors

Institutions

Publication Details

Journal
Biochemistry and Biophysics Reports
Published
2026-02-21
DOI
https://doi.org/10.1016/j.bbrep.2026.102509
Primary Topic
Protease and Inhibitor Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of a research use only (RUO) TIMP-1 assay for use with the automated Alinity i platform

Philip M. Hemken, Kathy S. Otis, John C. Prostko, Josie Corby et al.
Biochemistry and Biophysics Reports
Protease and Inhibitor Mechanisms
article

Evaluation of a research use only (RUO) TIMP-1 assay for use with the automated Alinity i platform

Philip M. Hemken, Kathy S. Otis, John C. Prostko, Josie Corby, Mekhi Baclig, Gina Pacenti, Joseph Munoz, Jose Ortiz, Jessica Grieshaber, Christine Rudolph, Mark Anderson
article en

Abstract

Tissue inhibitor of metalloproteinase-1 (TIMP-1) binds matrix metalloproteinases to regulate extracellular matrix turnover; elevated TIMP-1 has been associated with liver fibrosis, tumor progression, and fibrotic heart disease. We previously developed a quantitative, research use only (RUO) automated TIMP-1 immunoassay for use with the Abbott ARCHITECT <i>i</i> System; here, we describe the modification of the assay for use with the Abbott Alinity <i>i</i> system. The observed limit of detection was 1.42 ng/mL, and the calculated limit of quantitation was 2.44 ng/mL. In 100 normal plasma samples, TIMP-1 levels ranged from 106.23 ng/mL to 329.68 ng/mL. Deviation from linearity was ≤10% and precision was ≤10%CV across the analytical measuring interval from 2.44 to 500 ng/mL. No significant interference was observed for both conjugated and unconjugated bilirubin and hemoglobin at industry recommended test concentrations. Slight interference for protein and triglyceride resolved with titration to 12.15 g/dL and 750 mg/dL, respectively. Assay performance was optimal for plasma specimens in LiHep or EDTA. TIMP-1 in plasma samples was stable after multiple freeze-thaw cycles, up to 7 days at 2‒8 °C, and for up to 3 h on-board the Alinity <i>i</i> instrument. In cardiac plasma specimens, the assay consistently detected higher TIMP-1 in specimens with elevated troponin-I. The TIMP-1 immunoassay, modified for the Alinity<i>i</i> platform, detected TIMP-1 at physiologically relevant ranges. Future clinical development of the assay will focus on its clinical utility in various cardiac cohorts as a potential biomarker of fibrotic heart disease.

Biochemistry and Biophysics ReportsVol. 45
Discovery Laboratories (United States) (US)
Openalex Percentile: Top 46%
Protease and Inhibitor Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Evaluation of a research use only (RUO) TIMP-1 assay for use with the automated Alinity i platform — Philip M. Hemken, Kathy S. Otis, et al. · Biochemistry and Biophysics Reports (2026) | TGRS Research Map | TGRS