Influence of misfit direction and abutment type on torque loss and fracture strength of implant-supported fixed dental prostheses
Purpose: This study aimed to evaluate the effects of fatigue and misfit on the reverse torque, load-bearing capacity, and failure modes of cemented and screw-retained two-implant-supported fixed dental prostheses (FDPs). Materials and methods: Eighty CAD-CAM zirconia frameworks with porcelain veneering were assigned to three groups based on engaging (E) and non-engaging (N) abutment configurations: N-N, N-E, and E-E. Misfit scenarios included a control cast and 100 µm horizontal or vertical misfits. Frameworks were mounted on implants as screw-retained or cement-retained 3-unit FDPs, forming seven test groups and one control group (n = 10). Reverse torque was measured before and after thermomechanical cycling, followed by maximum failure load (Fmax) testing and failure mode analysis. Fmax was evaluated using two-way ANOVA, and reverse torque using mixed-effects ANOVA accounting for engagement, misfit, and implant position (α = 0.05). Conclusion: The engagement configuration of implants did not significantly affect reverse torque, fracture strength, or failure modes of cemented and screw-retained restorations, although horizontal misfit reduced load-bearing capacity.
Authors
- Ieva Gendvilienė (ORCID: https://orcid.org/0000-0002-1777-9781)
- Tan Fırat Eyüboğlu (ORCID: https://orcid.org/0000-0002-0308-9579)
- Ingrida Mischitz (ORCID: https://orcid.org/0000-0002-3628-2922)
- Daniel Kuleš (ORCID: https://orcid.org/0000-0002-2897-3213)
- Maisa Haiek (ORCID: https://orcid.org/0009-0008-9970-4222)
- Vygandas Rutkūnas
- Mutlu Özcan
- Liudas Auškalnis
Publication Details
- Journal
- Universität Zürich, ZORA
- Published
- 2025-12-01
- DOI
- https://doi.org/10.5167/uzh-292186
- Primary Topic
- Dental Implant Techniques and Outcomes
- Type
- article
- Field-Weighted Citation Impact
- 0.00