Influence of misfit direction and abutment type on torque loss and fracture strength of implant-supported fixed dental prostheses

Purpose: This study aimed to evaluate the effects of fatigue and misfit on the reverse torque, load-bearing capacity, and failure modes of cemented and screw-retained two-implant-supported fixed dental prostheses (FDPs). Materials and methods: Eighty CAD-CAM zirconia frameworks with porcelain veneering were assigned to three groups based on engaging (E) and non-engaging (N) abutment configurations: N-N, N-E, and E-E. Misfit scenarios included a control cast and 100 µm horizontal or vertical misfits. Frameworks were mounted on implants as screw-retained or cement-retained 3-unit FDPs, forming seven test groups and one control group (n = 10). Reverse torque was measured before and after thermomechanical cycling, followed by maximum failure load (Fmax) testing and failure mode analysis. Fmax was evaluated using two-way ANOVA, and reverse torque using mixed-effects ANOVA accounting for engagement, misfit, and implant position (α = 0.05). Conclusion: The engagement configuration of implants did not significantly affect reverse torque, fracture strength, or failure modes of cemented and screw-retained restorations, although horizontal misfit reduced load-bearing capacity.

Authors

Publication Details

Journal
Universität Zürich, ZORA
Published
2025-12-01
DOI
https://doi.org/10.5167/uzh-292186
Primary Topic
Dental Implant Techniques and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of misfit direction and abutment type on torque loss and fracture strength of implant-supported fixed dental prostheses

Ieva Gendvilienė, Tan Fırat Eyüboğlu, Ingrida Mischitz, Daniel Kuleš et al.
Universität Zürich, ZORA
Dental Implant Techniques and Outcomes
article

Influence of misfit direction and abutment type on torque loss and fracture strength of implant-supported fixed dental prostheses

Ieva Gendvilienė, Tan Fırat Eyüboğlu, Ingrida Mischitz, Daniel Kuleš, Maisa Haiek, Vygandas Rutkūnas, Mutlu Özcan, Liudas Auškalnis
article en

Abstract

Purpose: This study aimed to evaluate the effects of fatigue and misfit on the reverse torque, load-bearing capacity, and failure modes of cemented and screw-retained two-implant-supported fixed dental prostheses (FDPs). Materials and methods: Eighty CAD-CAM zirconia frameworks with porcelain veneering were assigned to three groups based on engaging (E) and non-engaging (N) abutment configurations: N-N, N-E, and E-E. Misfit scenarios included a control cast and 100 µm horizontal or vertical misfits. Frameworks were mounted on implants as screw-retained or cement-retained 3-unit FDPs, forming seven test groups and one control group (n = 10). Reverse torque was measured before and after thermomechanical cycling, followed by maximum failure load (Fmax) testing and failure mode analysis. Fmax was evaluated using two-way ANOVA, and reverse torque using mixed-effects ANOVA accounting for engagement, misfit, and implant position (α = 0.05). Conclusion: The engagement configuration of implants did not significantly affect reverse torque, fracture strength, or failure modes of cemented and screw-retained restorations, although horizontal misfit reduced load-bearing capacity.

Universität Zürich, ZORA
Sustainable cities and communities
Openalex Percentile: Top 23%
Dental Implant Techniques and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.