Clinical performance of a shielded diode for small field relative dosimetry in a 1.5 T MRI‐linac using measurements and simulations

This study reinforces the importance of detector choice for characterizing small fields in the presence of magnetic fields. The EDGE detector was shown to misrepresent the shape of the crossline profile for small fields in a 1.5T magnetic field. Film is recommended for determining the dose distribution shape, but it is difficult to position precisely and thus cannot be used to determine the lateral shift with high precision. The microDiamond detector was determined to most accurately characterize both the lateral shift and shape of the lateral dose distribution of the detectors investigated. The investigated ion chamber and shielded silicon diode should be used with caution for small fields in the presence of magnetic fields due to perturbances to profile shape.

Authors

Institutions

Publication Details

Journal
Medical Physics
Published
2026-02-01
DOI
https://doi.org/10.1002/mp.70356
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Clinical performance of a shielded diode for small field relative dosimetry in a 1.5 T MRI‐linac using measurements and simulations

William S. Ferris, Daniel Ellis Hyer, Damian Czarnecki, Joël St‐Aubin
Medical Physics
Advanced Radiotherapy Techniques
article

Clinical performance of a shielded diode for small field relative dosimetry in a 1.5 T MRI‐linac using measurements and simulations

William S. Ferris, Daniel Ellis Hyer, Damian Czarnecki, Joël St‐Aubin
article en

Abstract

This study reinforces the importance of detector choice for characterizing small fields in the presence of magnetic fields. The EDGE detector was shown to misrepresent the shape of the crossline profile for small fields in a 1.5T magnetic field. Film is recommended for determining the dose distribution shape, but it is difficult to position precisely and thus cannot be used to determine the lateral shift with high precision. The microDiamond detector was determined to most accurately characterize both the lateral shift and shape of the lateral dose distribution of the detectors investigated. The investigated ion chamber and shielded silicon diode should be used with caution for small fields in the presence of magnetic fields due to perturbances to profile shape.

Medical PhysicsVol. 53(2)
University of Iowa (US), Technische Hochschule Mittelhessen (DE)
Openalex Percentile: Top 44%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.