Vacuum-enhanced dielectric elastomer counterpulsation for in-vivo diastolic augmentation

Cardiac assist devices based on dielectric elastomer actuators (DEAs) offer a promising alternative to conventional mechanical circulatory support by providing soft and adaptive mechanical assistance. However, existing DEA-based systems face significant limitations in energy efficiency and effective hemodynamic modulation. This study presents a novel vacuum-enhanced DEA-based counterpulsation device and evaluates its performance in a porcine in-vivo model. We investigated two distinct implementation configurations: a direct attachment to the descending aorta and an integration into a bypass circuit mimicking an ascending aortic attachment. The device was actuated in synchronization with the cardiac cycle to provide real-time mechanical unloading. The results demonstrate that the vacuum-enhanced DEA system significantly improved diastolic augmentation (Mean: 39.1 ± 11.7%; Range: 29.2 to 55.0%) and achieved superior left ventricular unloading, marked by a remarkable reduction in end-diastolic pressure (Mean: −126.1 ± 65.6%; Range: −188.3 to −67.1%). Furthermore, the device reduced stroke work (Mean: −12.3 ± 14.0%; Range: −24.2 to 4.4%) and increased stroke volume (Mean: 3.9 ± 2.2%; Range: 2.5 to 7.1%) compared to baseline (device off). These findings establish vacuum-enhanced DEA actuation as a highly viable and impactful approach for next-generation cardiac assist devices, paving the way for further clinical optimization and application. • Evaluation of a vacuum-enhanced DEA for cardiac support in pigs. • Two configurations tested: direct aorta attachment or via a bypass circuit. • Device synchronized with heartbeat improved diastolic augmentation 39.1% ±11.7%. • Reduced end-diastolic pressure 126.1% ±65.6% and LV workload 12.3% ±14.0%. • Stroke volume rose by 3.9% ±2.2% from baseline. • Results show potential of device as a soft, adaptive cardiac assist system.

Authors

Institutions

Publication Details

Journal
Materials Today Advances
Published
2026-02-19
DOI
https://doi.org/10.1016/j.mtadv.2026.100730
Primary Topic
Dielectric materials and actuators
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Vacuum-enhanced dielectric elastomer counterpulsation for in-vivo diastolic augmentation

Jürgen Hörer, Yves Perriard, Francesco Clavica, Amine Benouhiba et al.
Materials Today Advances
Dielectric materials and actuators
article

Vacuum-enhanced dielectric elastomer counterpulsation for in-vivo diastolic augmentation

Jürgen Hörer, Yves Perriard, Francesco Clavica, Amine Benouhiba, Armando Walter, Silje Ekroll Jahren, Thierry Carrel, Dominik Obrist, Paul Philipp Heinisch, Yoan Civet
article en

Abstract

Cardiac assist devices based on dielectric elastomer actuators (DEAs) offer a promising alternative to conventional mechanical circulatory support by providing soft and adaptive mechanical assistance. However, existing DEA-based systems face significant limitations in energy efficiency and effective hemodynamic modulation. This study presents a novel vacuum-enhanced DEA-based counterpulsation device and evaluates its performance in a porcine in-vivo model. We investigated two distinct implementation configurations: a direct attachment to the descending aorta and an integration into a bypass circuit mimicking an ascending aortic attachment. The device was actuated in synchronization with the cardiac cycle to provide real-time mechanical unloading. The results demonstrate that the vacuum-enhanced DEA system significantly improved diastolic augmentation (Mean: 39.1 ± 11.7%; Range: 29.2 to 55.0%) and achieved superior left ventricular unloading, marked by a remarkable reduction in end-diastolic pressure (Mean: −126.1 ± 65.6%; Range: −188.3 to −67.1%). Furthermore, the device reduced stroke work (Mean: −12.3 ± 14.0%; Range: −24.2 to 4.4%) and increased stroke volume (Mean: 3.9 ± 2.2%; Range: 2.5 to 7.1%) compared to baseline (device off). These findings establish vacuum-enhanced DEA actuation as a highly viable and impactful approach for next-generation cardiac assist devices, paving the way for further clinical optimization and application. • Evaluation of a vacuum-enhanced DEA for cardiac support in pigs. • Two configurations tested: direct aorta attachment or via a bypass circuit. • Device synchronized with heartbeat improved diastolic augmentation 39.1% ±11.7%. • Reduced end-diastolic pressure 126.1% ±65.6% and LV workload 12.3% ±14.0%. • Stroke volume rose by 3.9% ±2.2% from baseline. • Results show potential of device as a soft, adaptive cardiac assist system.

Materials Today AdvancesVol. 29
University of Bern (CH), Bern University of Applied Sciences (CH), University of Zurich (CH), Deutsches Herzzentrum München (DE), École Polytechnique Fédérale de Lausanne (CH), Ludwig-Maximilians-Universität München (DE)
Affordable and clean energy
Openalex Percentile: Top 51%
Dielectric materials and actuators
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.