Development and validation of a 3D-printed dosimetry phantom for paediatric radiology

Radiation dosimetry is essential in the optimisation and justification of medical imaging procedures. However, the complexity of modern imaging equipment often surpasses the capabilities of standard dose calculation software, necessitating the use of commercially available dosimetry phantoms, which are often prohibitively expensive. This study aimed to develop a cost-effective, 3D-printed newborn-equivalent dosimetry phantom for measuring organ and whole-body effective doses. Several Polylactic Acid (PLA)-based filaments were investigated for tissue equivalency through Hounsfield-value analysis via micro-CT (40-70 kVp) and clinical CT (70-140 kVp) measurements. Standard PLA at 93% (ρ = 1.14 g/cm<sup>3</sup>) and 26% (ρ = 0.41 g/cm<sup>3</sup>) infill density was selected for soft tissue and lung, respectively, while StoneFil composite PLA (FormFutura) at 81% (ρ = 1.21 g/cm<sup>3</sup>) infill was chosen for bone. The phantom was modelled on a modified Cristy and Eckerman newborn design, with 21 sections generated using MATLAB and printed on a Bambu Lab X1 Carbon 3D printer. A total of 186 thermoluminescent dosimeter (TLD) capsules were embedded in the phantom, and TLD measurements from whole-body 60 kVp radiographs were compared with Monte Carlo (PCXMC 2.0) simulations for validation. The phantom demonstrated accurate dosimetry for the radiographic exposure, with average organ doses closely matching the simulated exposure, and the effective dose (ICRP 103) within 2% of the simulation. The phantom required 135 h to print, with a material cost of A$165. This study successfully developed and validated a cost-effective dosimetry phantom for paediatric radiography, with the potential to print larger phantoms for older children. Future work will explore the phantom's application in other X-ray imaging modalities.

Authors

Institutions

Publication Details

Journal
Physical and Engineering Sciences in Medicine
Published
2026-02-18
DOI
https://doi.org/10.1007/s13246-026-01698-3
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development and validation of a 3D-printed dosimetry phantom for paediatric radiology

C. L. Williams, Rikki Nezich, Jonathon Stone, Adriano Polpo et al.
Physical and Engineering Sciences in Medicine
Radiation Dose and Imaging
article

Development and validation of a 3D-printed dosimetry phantom for paediatric radiology

C. L. Williams, Rikki Nezich, Jonathon Stone, Adriano Polpo, Pejman Rowshanfarzad
article en

Abstract

Radiation dosimetry is essential in the optimisation and justification of medical imaging procedures. However, the complexity of modern imaging equipment often surpasses the capabilities of standard dose calculation software, necessitating the use of commercially available dosimetry phantoms, which are often prohibitively expensive. This study aimed to develop a cost-effective, 3D-printed newborn-equivalent dosimetry phantom for measuring organ and whole-body effective doses. Several Polylactic Acid (PLA)-based filaments were investigated for tissue equivalency through Hounsfield-value analysis via micro-CT (40-70 kVp) and clinical CT (70-140 kVp) measurements. Standard PLA at 93% (ρ = 1.14 g/cm<sup>3</sup>) and 26% (ρ = 0.41 g/cm<sup>3</sup>) infill density was selected for soft tissue and lung, respectively, while StoneFil composite PLA (FormFutura) at 81% (ρ = 1.21 g/cm<sup>3</sup>) infill was chosen for bone. The phantom was modelled on a modified Cristy and Eckerman newborn design, with 21 sections generated using MATLAB and printed on a Bambu Lab X1 Carbon 3D printer. A total of 186 thermoluminescent dosimeter (TLD) capsules were embedded in the phantom, and TLD measurements from whole-body 60 kVp radiographs were compared with Monte Carlo (PCXMC 2.0) simulations for validation. The phantom demonstrated accurate dosimetry for the radiographic exposure, with average organ doses closely matching the simulated exposure, and the effective dose (ICRP 103) within 2% of the simulation. The phantom required 135 h to print, with a material cost of A$165. This study successfully developed and validated a cost-effective dosimetry phantom for paediatric radiography, with the potential to print larger phantoms for older children. Future work will explore the phantom's application in other X-ray imaging modalities.

Physical and Engineering Sciences in Medicine
The University of Western Australia (AU), Princess Margaret Hospital for Children (AU), Sir Charles Gairdner Hospital (AU), Act Health (AU), Perth Children's Hospital
Openalex Percentile: Top 51%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.