Evaluation of the InTempo path set for CyberKnife prostate and lung SBRT: A single‐institution experience
InTempo-compatible path sets do not significantly compromise plan quality for prostate or lung SBRT, provided adequate node availability. Specifically, the Prostate_InTempo and Prostate_Short path sets demonstrated a reduction in delivery time and an increase in adaptive imaging frequency compared with the Prostate path set. However, the Reduced_Prostate and Reduced_Prostate_InTempo result in inferior plan quality and reduced deliverability and should be used with caution. These findings support the selective use of InTempo imaging in SBRT planning without sacrificing dosimetric integrity.
Authors
- Shari Rudoler
- Y. Vinogradskiy
- Eric Gressen (ORCID: https://orcid.org/0000-0002-8243-5459)
- Yingcui Jia
- Lei Fu
- Qianyi Xu
Institutions
- Thomas Jefferson University (US)
Publication Details
- Journal
- Journal of Applied Clinical Medical Physics
- Published
- 2026-02-01
- DOI
- https://doi.org/10.1002/acm2.70509
- Primary Topic
- Advanced Radiotherapy Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00