Comparative Buckling Strength and Metallurgical Analysis of Five Classic NiTi Endodontic Rotary Files

Objective: Nickel-titanium (NiTi) rotary instruments have revolutionized endodontic practice through continuous advancements in metallurgy and design. Despite these improvements, mechanical failure remains a clinical concern. This study aimed to evaluate and compare the design features, metallurgical properties, and buckling resistance of five widely used NiTi rotary endodontic systems.b Methods: A total of 250 new NiTi rotary instruments from fi e systems (ProTaper Next, Mtwo, ProFile, EndoSequence, and GT Series X) were analyzed. Design features were assessed using dental microscopy and scanning electron microscopy. Metallurgical properties were evaluated through energy-dispersive X-ray spectroscopy and differential scanning calorimetry. Buckling resistance was measured using a universal testing machine equipped with a 1 kN load cell, applying a compressive load at 1 mm/min until 1 mm of lateral displacement was achieved. Statistical analysis was performed using the Shapiro–Wilk test to assess normality, followed by the non-parametric Kruskal-Wallis test to compare groups. A signifi e level of p<0.05 was adopted. Results: ProFile instruments exhibited the highest number of spirals (19) and spiral density (1.19 spirals/mm), while GT Series X featured the shortest cutting blade length (≤12 mm). All systems demonstrated near-equiatomic nickeltitanium ratios. ProTaper Next and GT Series X showed higher R-phase and austenitic transformation temperatures. Buckling resistance was significantly greater in the ProFile (0.04 and 0.06 taper) and EndoSequence 35/.06 and 40/.06 instruments (p<0.05). In contrast, EndoSequence 0.04 files, Mtwo, and ProTaper Next exhibited lower resistance. Conclusion: Design features, taper, and metallurgical composition significantly influence the buckling resistance of NiTi rotary endodontic instruments.

Authors

Publication Details

Journal
Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Published
2025-09-09
Primary Topic
Endodontics and Root Canal Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparative Buckling Strength and Metallurgical Analysis of Five Classic NiTi Endodontic Rotary Files

J. Estêvão Martins, Francisco Manuel Braz Fernandes, Abayomi Baruwa, João Caramês
Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Endodontics and Root Canal Treatments
article

Comparative Buckling Strength and Metallurgical Analysis of Five Classic NiTi Endodontic Rotary Files

J. Estêvão Martins, Francisco Manuel Braz Fernandes, Abayomi Baruwa, João Caramês
article en

Abstract

Objective: Nickel-titanium (NiTi) rotary instruments have revolutionized endodontic practice through continuous advancements in metallurgy and design. Despite these improvements, mechanical failure remains a clinical concern. This study aimed to evaluate and compare the design features, metallurgical properties, and buckling resistance of five widely used NiTi rotary endodontic systems.b Methods: A total of 250 new NiTi rotary instruments from fi e systems (ProTaper Next, Mtwo, ProFile, EndoSequence, and GT Series X) were analyzed. Design features were assessed using dental microscopy and scanning electron microscopy. Metallurgical properties were evaluated through energy-dispersive X-ray spectroscopy and differential scanning calorimetry. Buckling resistance was measured using a universal testing machine equipped with a 1 kN load cell, applying a compressive load at 1 mm/min until 1 mm of lateral displacement was achieved. Statistical analysis was performed using the Shapiro–Wilk test to assess normality, followed by the non-parametric Kruskal-Wallis test to compare groups. A signifi e level of p<0.05 was adopted. Results: ProFile instruments exhibited the highest number of spirals (19) and spiral density (1.19 spirals/mm), while GT Series X featured the shortest cutting blade length (≤12 mm). All systems demonstrated near-equiatomic nickeltitanium ratios. ProTaper Next and GT Series X showed higher R-phase and austenitic transformation temperatures. Buckling resistance was significantly greater in the ProFile (0.04 and 0.06 taper) and EndoSequence 35/.06 and 40/.06 instruments (p<0.05). In contrast, EndoSequence 0.04 files, Mtwo, and ProTaper Next exhibited lower resistance. Conclusion: Design features, taper, and metallurgical composition significantly influence the buckling resistance of NiTi rotary endodontic instruments.

Portuguese National Funding Agency for Science, Research and Technology (RCAAP Project by FCT)
Affordable and clean energy
Openalex Percentile: Top 23%
Endodontics and Root Canal Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.