ABSTRACT In this study, sum peaks related to the Ca Kα line (3.69 keV) were observed in total reflection X‐ray fluorescence spectra of dry residues of a solution sample containing a large amount of Ca, which were measured with a silicon drift detector. When dry residues containing 1 μg of Ca were measured, a peak that would be the sum peak of the Si Kα line (1.74 keV) originating from sample holders and the Ca Kα line appeared at around 5.4 keV in each spectrum. Net intensities of the sum peak, which can overlap with the Cr Kα line (5.41 keV), were comparable to that of the Cr Kα line from the dry residue containing 200 pg of Cr. The sum peak can be misinterpreted as the Cr Kα line, and this misinterpretation leads to a false decision that a trace amount of Cr is contained in the dry residue of a water sample that does not actually contain Cr.

Authors

Institutions

Publication Details

Journal
X-Ray Spectrometry
Published
2026-02-17
DOI
https://doi.org/10.1002/xrs.70086
Primary Topic
X-ray Spectroscopy and Fluorescence Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sum Peaks in a Total Reflection X‐Ray Fluorescence Spectrum of the Dry Residue of a Solution Sample Containing Ca

Yusaku Oki, Shinsuke Kunimura, Kosei Oikawa, Masahide Wada et al.
X-Ray Spectrometry
X-ray Spectroscopy and Fluorescence Analysis
article

Sum Peaks in a Total Reflection X‐Ray Fluorescence Spectrum of the Dry Residue of a Solution Sample Containing Ca

Yusaku Oki, Shinsuke Kunimura, Kosei Oikawa, Masahide Wada, Yume Kubota
article en

Abstract

ABSTRACT In this study, sum peaks related to the Ca Kα line (3.69 keV) were observed in total reflection X‐ray fluorescence spectra of dry residues of a solution sample containing a large amount of Ca, which were measured with a silicon drift detector. When dry residues containing 1 μg of Ca were measured, a peak that would be the sum peak of the Si Kα line (1.74 keV) originating from sample holders and the Ca Kα line appeared at around 5.4 keV in each spectrum. Net intensities of the sum peak, which can overlap with the Cr Kα line (5.41 keV), were comparable to that of the Cr Kα line from the dry residue containing 200 pg of Cr. The sum peak can be misinterpreted as the Cr Kα line, and this misinterpretation leads to a false decision that a trace amount of Cr is contained in the dry residue of a water sample that does not actually contain Cr.

X-Ray Spectrometry
Tokyo University of Science (JP), Advanced Institute of Industrial Technology (JP)
Clean water and sanitation
Openalex Percentile: Top 46%
X-ray Spectroscopy and Fluorescence Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Sum Peaks in a Total Reflection X‐Ray Fluorescence Spectrum of the Dry Residue of a Solution Sample Containing Ca — Yusaku Oki, Shinsuke Kunimura, et al. · X-Ray Spectrometry (2026) | TGRS Research Map | TGRS