Multi‐institutional validation of hypersight CBCT‐based dose calculation on O‐ring linacs
HyperSight CBCT provides accurate dose calculations when properly calibrated. Phantom-based validation demonstrated sub-2% deviations and strong agreement with CT, supporting its clinical use in adaptive radiotherapy.