Evaluation of sample integrity and turnaround time upon implementation of a high-speed carrier-free pneumatic tube system

BackgroundPneumatic tube systems (PTSs) have revolutionized sample transport efficiency, yet concerns about sample integrity remain unresolved. This study evaluated the AirLab system (PazKorea, Hanam-si, Korea), a novel high-speed carrier-free PTS, for its impact on sample integrity and turnaround time (TAT) under controlled pre-analytical conditions.MethodsBlood specimens were collected in duplicate from 42 participants (30 healthy volunteers and 12 outpatients with chronic alcohol use) and randomly assigned to PTS or manual transport. Multiple laboratory parameters were analysed, including chemistry, hematology, coagulation, and heavy metal tests. Sample integrity was assessed by comparing test results between transport methods using both statistical analysis and clinical significance criteria based on total allowable error (TAE). Pre-analytical, analytical and total TATs were compared between transport methods.ResultsAmong 20 measurands, four showed statistically significant differences: glucose, potassium (K), lactate dehydrogenase (LD), and mean corpuscular volume (MCV). PTS yielded higher results for glucose, K and LD (average percentage biases: 1.88%, 2.01%, and 15.80%, respectively), while producing lower results for MCV (average percentage bias: -0.47%). Only LD showed clinically significant differences, as its average percentage bias exceeded its desirable TAE. PTS significantly reduced both pre-analytical and total TATs for all test types, with reductions ranging from 12.12% to 22.82% and 7.32% to 8.98%, respectively, whereas analytical TATs remained unchanged.ConclusionsThe AirLab system effectively preserves sample integrity for most laboratory parameters while improving operational efficiency through TAT reduction. However, the clinically significant elevation in LD levels necessitates careful interpretation when employing this system in clinical laboratories.

Authors

Institutions

Publication Details

Journal
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Published
2026-02-16
DOI
https://doi.org/10.1177/00045632261428924
Primary Topic
Clinical Laboratory Practices and Quality Control
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of sample integrity and turnaround time upon implementation of a high-speed carrier-free pneumatic tube system

Soo-Youn Lee Lee, Hyun-Young Kim, Hyung-Doo Park, Hyun-Woo Lee et al.
Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Clinical Laboratory Practices and Quality Control
article

Evaluation of sample integrity and turnaround time upon implementation of a high-speed carrier-free pneumatic tube system

Soo-Youn Lee Lee, Hyun-Young Kim, Hyung-Doo Park, Hyun-Woo Lee, Sang-Mi Kim
article en

Abstract

BackgroundPneumatic tube systems (PTSs) have revolutionized sample transport efficiency, yet concerns about sample integrity remain unresolved. This study evaluated the AirLab system (PazKorea, Hanam-si, Korea), a novel high-speed carrier-free PTS, for its impact on sample integrity and turnaround time (TAT) under controlled pre-analytical conditions.MethodsBlood specimens were collected in duplicate from 42 participants (30 healthy volunteers and 12 outpatients with chronic alcohol use) and randomly assigned to PTS or manual transport. Multiple laboratory parameters were analysed, including chemistry, hematology, coagulation, and heavy metal tests. Sample integrity was assessed by comparing test results between transport methods using both statistical analysis and clinical significance criteria based on total allowable error (TAE). Pre-analytical, analytical and total TATs were compared between transport methods.ResultsAmong 20 measurands, four showed statistically significant differences: glucose, potassium (K), lactate dehydrogenase (LD), and mean corpuscular volume (MCV). PTS yielded higher results for glucose, K and LD (average percentage biases: 1.88%, 2.01%, and 15.80%, respectively), while producing lower results for MCV (average percentage bias: -0.47%). Only LD showed clinically significant differences, as its average percentage bias exceeded its desirable TAE. PTS significantly reduced both pre-analytical and total TATs for all test types, with reductions ranging from 12.12% to 22.82% and 7.32% to 8.98%, respectively, whereas analytical TATs remained unchanged.ConclusionsThe AirLab system effectively preserves sample integrity for most laboratory parameters while improving operational efficiency through TAT reduction. However, the clinically significant elevation in LD levels necessitates careful interpretation when employing this system in clinical laboratories.

Annals of Clinical Biochemistry International Journal of Laboratory Medicine
Samsung Medical Center (KR), Chosun University Hospital (KR), Sungkyunkwan University (KR)
Good health and well-being
Openalex Percentile: Top 51%
Clinical Laboratory Practices and Quality Control
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.