Inner Branch Off the Shelf Technology for Chronic Post-dissection Thoraco-Abdominal Aneurysm with Narrow True Lumen: Results of a European Multicentre Study.

Objective This study aimed to evaluate the early and midterm outcomes of endovascular repair of aortic dissection (AD) using a new pre-cannulated, inner branch based, off the shelf stent graft (E-nside; JOTEC, CryoLife, Inc.).Methods Data from an international multicentre registry on patients treated with the E-nside endograft for AD were analysed. Pre-operative clinical and anatomical characteristics, procedural data, and early and follow up outcomes were recorded in a dedicated electronic data capture system. The primary endpoint was technical success. Secondary endpoints were overall death, aortic related death, aortic related re-intervention rate, endoleak rate, and target vessel instability (TVI) at 30 days and during follow up.Results Thirty four patients treated in 14 high volume European institutions between January 2020 and May 2024 were included. The mean patient age ± standard deviation was 66 ± 9.8 years, and 30 (88%) were male. The mean true lumen aortic diameter at the visceral aorta was 24.5 ± 7.7 mm. The median procedure time was 300 minutes (interquartile range 215, 373). The technical success rate was 97%, due to one case of placement of the coeliac trunk covered stent into the branch of the superior mesenteric artery and vice versa. The 30 day mortality and re-intervention rates were 6% (n = 2) and 9% (n = 3), respectively. At a mean follow up of 15.1 ± 7.5 months, there were no aortic related deaths. Freedom from aortic related death and all cause death at 12 months was 97% and 94%, respectively. The TVI rate during the follow up was 13% (four of 32). The total re-intervention rate was 16% (five of 32).Conclusion Endovascular treatment of AD with an off the shelf, inner branched endograft is feasible, with acceptable short and midterm outcomes. The results showed good technical implantation safety and efficacy. Long term follow up results are needed to better define the clinical role of this novel endograft.

Authors

Institutions

Publication Details

Journal
Open Access CRIS of the University of Bern
Published
2025-10-01
DOI
https://doi.org/10.48620/94660
Primary Topic
Aortic Disease and Treatment Approaches
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Inner Branch Off the Shelf Technology for Chronic Post-dissection Thoraco-Abdominal Aneurysm with Narrow True Lumen: Results of a European Multicentre Study.

Giovanni Luigi Bruno. Corrado Cariddi Valentina Tinelli, Vladimir Makaloski, Michele Piazza, Mario Lescan et al.
Open Access CRIS of the University of Bern
Aortic Disease and Treatment Approaches
article

Inner Branch Off the Shelf Technology for Chronic Post-dissection Thoraco-Abdominal Aneurysm with Narrow True Lumen: Results of a European Multicentre Study.

Giovanni Luigi Bruno. Corrado Cariddi Valentina Tinelli, Vladimir Makaloski, Michele Piazza, Mario Lescan, Simona Sica, Fabio Verzini, Luca Bertoglio, Theodosios Bisdas, Yamume Tshomba, Giovanni Pratesi
article en

Abstract

Objective This study aimed to evaluate the early and midterm outcomes of endovascular repair of aortic dissection (AD) using a new pre-cannulated, inner branch based, off the shelf stent graft (E-nside; JOTEC, CryoLife, Inc.).Methods Data from an international multicentre registry on patients treated with the E-nside endograft for AD were analysed. Pre-operative clinical and anatomical characteristics, procedural data, and early and follow up outcomes were recorded in a dedicated electronic data capture system. The primary endpoint was technical success. Secondary endpoints were overall death, aortic related death, aortic related re-intervention rate, endoleak rate, and target vessel instability (TVI) at 30 days and during follow up.Results Thirty four patients treated in 14 high volume European institutions between January 2020 and May 2024 were included. The mean patient age ± standard deviation was 66 ± 9.8 years, and 30 (88%) were male. The mean true lumen aortic diameter at the visceral aorta was 24.5 ± 7.7 mm. The median procedure time was 300 minutes (interquartile range 215, 373). The technical success rate was 97%, due to one case of placement of the coeliac trunk covered stent into the branch of the superior mesenteric artery and vice versa. The 30 day mortality and re-intervention rates were 6% (n = 2) and 9% (n = 3), respectively. At a mean follow up of 15.1 ± 7.5 months, there were no aortic related deaths. Freedom from aortic related death and all cause death at 12 months was 97% and 94%, respectively. The TVI rate during the follow up was 13% (four of 32). The total re-intervention rate was 16% (five of 32).Conclusion Endovascular treatment of AD with an off the shelf, inner branched endograft is feasible, with acceptable short and midterm outcomes. The results showed good technical implantation safety and efficacy. Long term follow up results are needed to better define the clinical role of this novel endograft.

Open Access CRIS of the University of Bern
United Heart and Vascular Clinic (US), Center for Vascular Biology Research (US)
Good health and well-being
Openalex Percentile: Top 24%
Aortic Disease and Treatment Approaches
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.